1
1.0Object
1.1
To test the performance of Rossmax NE100 nebulizer
2.0Equipment List
2.1Rossmax NE100 Nebulizer*2
2.2Rossmax Nebulizer kit*2
2.3Malvern Spraytec particle size analyer
2.4Marple 298 Cascade Impactor
2.5Chroma 61602 Programmable AC Source
2.6Shimadzu AUW120D microbalance
2.7A.P. Buck, Inc. Libra Plus LP-5 sampling pump
2.8SSI P51-6BarS-A-MD-20mA pressure meter
2.9Golden Mountain Enterprise Co. Ltd. F33L0096 flow meter
2.10 Humidity/Temperature Meter
2.11 Taiwan Biotech Co., Ltd 0.9% Saline solution
2.12 2.5% NaF solution
2.13 Atrovent Ipratropium Bromide
2.14
Atrovent Flixotide
2.15 AstraZeneca Terbutaline Sulphate
2.16 Ventoline (2.5mg) Salbutamol/Sulphate
2.17 Casio Timer
3.0Testing Items
3.1Aerosol Particle Size Distribution Testing(By Malvern Spraytec)
3.2Aerosol Particle Size Distribution Testing(By Marple 298 Cascade Impactor)
3.3Nebulization Rate Testing(Including drugs testing)
3.4Residual Volume Testing
3.5Reliability Test
4.0Testing Procedure
4.1Aerosol Particle Size Distribution Testing(By Malvern Spraytec)
4.1.1Each sample should be tested with 2.5ml 0.9% saline solution for 3 minutes.
4.1.2Add 2.5ml 0.9% saline solution into the nebulizer kit,
4.1.3Connect the nebulizer kit with NE100 and put at the testing site, the nebulizer
kit’s outlet must be kept at 3.0 cm from the laser beam.
4.1.4Start recording Spraytec for more than 15 secs, then start NE100 for testing.
4.1.5After 3.0 minutes have been reached, stop the NE100 and then stop Spraytec.
4.1.6Checks Spraytec records
Fig 1. Malvern Spraytec and Testing site