Bekijk gratis de handleiding van PCE Instruments PCE-RVP 2200 (5 pagina’s), behorend tot de categorie Meetapparatuur. Deze gids werd als nuttig beoordeeld door 15 mensen en kreeg gemiddeld 4.9 sterren uit 8 reviews. Heb je een vraag over PCE Instruments PCE-RVP 2200 of wil je andere gebruikers van dit product iets vragen? Stel een vraag
Pagina 1/5

User Manualting
Manual
Vacuum Pump PCE-RVP 180 / PCE-RVP 2200
Vacuum Pump PCE-RVP 180 / PCE-RVP 2200
Vacuum Pump PCE-RVP 180 / PCE-RVP 2200
Vacuum Pump PCE-RVP 180 / PCE-RVP 2200Vacuum Pump PCE-RVP 180 / PCE-RVP 2200

01 02
A
Structure and Technical Data
Structure
Oil Tank
Oil Filling Cap
Handle
Fan Cover
Motor
Base
Technical Data
Applications
PCE-RVP vacuum pumps are d used to create a vacuum by evacuating gas from a seale
vessel. The pumps are applicable for refrigeration maintenance to charge refrigerants of
CFC, HCFH and HFC such as R12/R22/R134a and air extraction in printing machines, food
processing, gas analysis, thermoplastic molding, etc.. In some cases they can be used as fore
pumps for high vacuum equipment.
Oil Sight Glass
BApplications and Features
Oil Drain Plug
Model
Rated power supply
Ultim ate Vac uum (M icron )
Motor Power (HP)
Intake Fittings
Dimensions (mm)
Oil Capacity (ml)
N.W.(kg)
Single Stage Rotary Vane Vacuum Pump
1/4" & 3/8" Flare
220V~50Hz/110V~60Hz
PCE-RVP 180
8/9.6
226/270
150
3/4
780
14
383×138×255
Flow Rate CFM
L/min
PCE-RVP 2200
12/14.4
340/407
15
1 1/2
620
15.7
383×138×255
220V~50Hz/110V~60Hz
Double Stage Rotary Vane Vacuum Pump
Model
Rated power supply
Ultim ate Vac uum (M icron )
Motor Power (HP)
Intake Fittings
Dimensions (mm)
Oil Capacity (ml)
N.W. (kg)
Flow Rate CFM
L/min
1/4" & 3/8" Flare

03 04
Features
1. High ultimate vacuum and high pumping speed
High ultimate vacuum and high pumping speed are achieved by taking integral cylinder
construction with high precision.
2. High efficient filtration
High quality filter element is used to reduce oil mist effectively.
3. Comfortable and robust handle
Soft rubber lined handle, comfortable to carry and use.
4. Elegant appearance
Unique and elegant appearance with square-shaped pump casing and fan cover.
CPreparation, Operation and Safety Precautions
Preparation
1. Make sure the power supply matches the voltage and frequency marked on the pump
nameplate.
2. Make sure the switch is in the OFF position before connecting electrical power to the pump.
3. Fill oil till the oil level reaches the middle position on oil sight glass. Low oil level will cause
pump performance loss while high oil level may result in oil spraying.
Operation
1. Before using the vacuum pump, remove the inlet cap and connect the pump to the sealed
vessel (Check and make sure the thread of the pipe fits that of the inlet port). The length of the
connecting pipe shall be as short as possible.
2. Check and make sure the air inlet is well sealed without any leakage.
3. Switch on the pump to start operation.
4. Close the valve between the pump and the sealed vessel when the operation is finished.
5. Switch off the pump and remove the power plug.
6. Remove the connecting pipe.
Safety Precautions
Always wear goggles when handling air-conditioning refrigerant.
Do not contact the refrigerant directly! Personal injury may occur.
Hot surface! Do not touch the oil tank or pump casing during pump operation.
Risk of electrical shock! Make sure all equipment are properly grounded when connecting
the power supply.
The pump is not allowed to operate more than 3 minutes while the inlet is exposed to the
atmosphere.
The ambient temperature for proper operation is 5°C - 40°C.
The voltage of the pump is 220V±10%/50Hz. The power socket must be grounded.
The pump body will be damaged to take out the refrigerant when the pump chamber is
pressurized. It is recommended to use special equipment to do this work.
WARNING!
To avoid personal injury, please carefully read and follow the instructions of this manual.
Product specificaties
Merk: | PCE Instruments |
Categorie: | Meetapparatuur |
Model: | PCE-RVP 2200 |
Heb je hulp nodig?
Als je hulp nodig hebt met PCE Instruments PCE-RVP 2200 stel dan hieronder een vraag en andere gebruikers zullen je antwoorden
Handleiding Meetapparatuur PCE Instruments
16 Juni 2025
16 Juni 2025
15 Juni 2025
15 Juni 2025
25 Mei 2025
22 Mei 2025
22 Mei 2025
22 Mei 2025
14 Mei 2025
14 Mei 2025
Handleiding Meetapparatuur
- Gude
- Mitsubishi
- Metrix
- IHealth
- Greisinger
- Hama
- Aeroqual
- Apollo Ultrasonic
- Beckmann & Egle
- NEO Tools
- Silverline
- Arya
- Honeywell
- Maximum
- Nedis
Nieuwste handleidingen voor Meetapparatuur
16 Juni 2025
15 Juni 2025
15 Juni 2025
15 Juni 2025
15 Juni 2025
15 Juni 2025
15 Juni 2025
15 Juni 2025
15 Juni 2025
12 Juni 2025