Instructions for Use
Before operating the unit, please read this manual thoroughly
and retain it for future reference.
UP-991AD/UP-971AD
© 2014 Sony Corporation
Hybrid Graphic Printer
4-488-657-18 (1)
2023-06
2
Indications for Use/Intended Use
The Sony UP-971AD and UP-991AD are compact,
medical grade black and white printers that can accept
both analog and digital signal inputs. They are designed
to be integrated into radiology imaging systems such as
mobile c-arm, ultrasound, cardiac cath lab and other
compatible medical imaging systems and produce hard
copy prints of still images captured by these systems for
the patient record or for referrals.
Notes
• Output images from this equipment cannot be used for
diagnostic use.
• This equipment is for medical professionals.
• This equipment is intended for use in medical
environments, such as clinics, examination rooms,
and operating rooms.
To reduce the risk of fire or electric shock, do
not expose this equipment to rain or moisture.
To avoid electrical shock, do not open the
cabinet. Refer servicing to qualified personnel
only.
No modification of this equipment is allowed.
WARNING
To avoid the risk of electric shock, this
equipment must only be connected to a supply
mains with protective earth.
WARNING
To disconnect the main power, unplug the power plug.
When installing the unit, incorporate a readily
accessible disconnect device in the fixed wiring, or
connect the power plug to an easily accessible socket-
outlet near the unit.
Do not position the ME equipment where it is difficult to
unplug the power plug.
If a fault should occur during operation of the unit,
operate the disconnect device to switch the power
supply off, or disconnect the power plug.
c
WARNING
Symbols on the product
Consult the instructions for use
Follow the directions in the instructions for
use for parts of the unit on which this symbol
appears.
This symbol indicates the manufacturer, and
appears next to the manufacturer’s name and
address.
This symbol indicates the date of
manufacture.
This symbol indicates the serial number.
This symbol indicates the Unique Device
Identifier (UDI), and appears next to the bar
code representation of the Unique Device
Identification.
This symbol indicates the equipotential
terminal which brings the various parts of a
system to the same potential.
Storage and transport temperature
This symbol indicates the acceptable
temperature range for storage and transport
environments.
Storage and transport humidity
This symbol indicates the acceptable
humidity range for storage and transport
environments.
Storage and transport pressure
This symbol indicates the acceptable
atmospheric pressure range for storage and
transport environments.
This symbol indicates the Importer, and
appears next to the Importer’s name and
registered office address.
This symbol indicates the European
Community representative, and appears next
to the European Community representative’s
name and address.
This symbol indicates the UK Responsible
Person, and appears next to the UK
Responsible Person’s name and address.
This symbol indicates the Swiss authorized
representative, and appears next to the Swiss
authorized representative’s name and
address.
This symbol indicates the medical device in
the European Community.
3
For customers in the U.S.A.
This equipment has been tested and found to comply
with the limits for a Class A digital device, pursuant to
part 15 of the FCC Rules. These limits are designed to
provide reasonable protection against harmful
interference when the equipment is operated in a
commercial environment. This equipment generates,
uses and can radiate radio frequency energy and, if not
installed and used in accordance with the instruction
manual, may cause harmful interference to radio
communications. Operation of this equipment in a
residential area is likely to cause harmful interference in
which case the user will be required to correct the
interference at his own expense.
You are cautioned that any changes or modifications not
expressly approved in this manual could void your
authority to operate this equipment.
All interface cables used to connect peripherals must be
shielded in order to comply with the limits for a digital
device pursuant to Subpart B of part 15 of FCC Rules.
For customers in the U.S.A.
Caution
Federal law (United States of America) restricts this
device to sale by or on the order of a licensed healthcare
practitioner.
For customers in Canada
This unit has been certified according to Standard CAN/
CSA-C22.2 No.60601-1.
Important safeguards and notices for use
in the medical environments
1. All devices connected to the unit must be certified or
compliant according to IEC 60601-1, IEC 60950-1,
IEC 62368-1 and IEC 60065 standards and other
IEC/ISO standards applicable to the devices.
2. Furthermore, the system as a whole must comply
with IEC 60601-1 standards. All peripheral devices
connected to the signal input/output sections of the
unit constitute the medical-use system, and
therefore, the user is responsible for ensuring that the
system as a whole complies with IEC 60601-1
standards. If in doubt, consult qualified Sony service
personnel.
3. Connecting the unit to other devices may increase the
leakage current.
4. For all peripheral devices connected to the unit that
operate on commercial power supplies and do not
comply with IEC 60601-1 standards, incorporate an
isolation transformer that complies with IEC 60601-
1 standards and connect to the commercial power
supply via the transformer.
5. The unit generates, uses, and may radiate radio
frequency energy. If it is not installed and used in
accordance with the instruction manual, it may cause
interference on other devices. If the unit causes
interference (which can be determined by
disconnecting the power cord from the unit), try the
following.
– Relocate the unit with respect to the affected
devices.
– Connect the unit and the affected devices to
different branch circuits.
For more information, consult qualified Sony service
personnel.
(Applicable standard: IEC 60601-1-2)
This symbol indicates sharp edges.
For accessories
This symbol indicates the batch code.
This symbol indicates “Do not reuse”.


Product specificaties

Merk: Sony
Categorie: Printer
Model: UP-971AD
Soort bediening: Buttons, Rotary
Kleur van het product: Wit
Ingebouwd display: Ja
Gewicht: 7000 g
Breedte: 316 mm
Diepte: 265 mm
Hoogte: 132.5 mm
Stroombron: AC
Kleur: Nee
Resolutie: 162.5 lpi
Aantal USB 2.0-poorten: 1
Formaat analoog signaal: NTSC, PAL
Temperatuur bij opslag: -20 - 60 °C
Luchtvochtigheid bij opslag: 20 - 80 procent
Printkleuren: Zwart
Printtechnologie: Direct thermisch printen
Grijsniveaus: 256
BNC input ports: 1
Maximale print breedte: 210 mm
AC-ingangsspanning: 100-240 V
AC-ingangsfrequentie: 50 - 60 Hz
Bedrijfstemperatuur (T-T): 5 - 35 °C
Relatieve vochtigheid in bedrijf (V-V): 20 - 80 procent
Afdruktijd (min): 8 s
Gradaties: 8 bits
Ingangsstroom (max): 2.9 A
Ingangsstroom (min): 1.2 A