Features
Model
T90 T110VDE T130VDE T150VDE
Complies with EN 61243-3:2014 • •
LED Indication Range: 12 V to 690 V dc and ac • •
V Display: Multiple LED Bargraph • •
Independent ELV indicator LED, indicates if >50 V ac/120 V dc is
present even in the event of no battery power or main circuit failure • •
LCD Indication Range: 6 V to 690 V dc and ac • •
V Display: Digital LCD 3½ digit (1 V resolution) • •
Resistance Measurement: LCD 3½ digit (0 to 1999 Ω/1Ω resolution)
LCD Backlight • •
Display HOLD: Freeze/unfreeze display with voltage or resistance
measurement • •
CAT II 690 V / CAT III 600 V
CAT III 690 V / CAT IV 600 V • •
Rugged, Double-Insulated Wire • •
Fixed Impedance ~200 kΩ ( 3.5 mA @ 690 V) • •
Switchable Load by 2 pushbuttons (~30 mA @ 230 V) • •
Vibration During Load (when 2 switchable load pushbuttons are
pushed)
• •
Single-Pole Phase Test (also operates with gloves) • •
Rotary Field Direction (also operates with gloves) • •
Continuity Test / Diode Test • •
Torch • •
Beeper for Continuity/Phase/ACV • •
IP54
IP64 • •
Slim Metal Probe Tips (threaded base for included tip accessories) • •
Probe Tip Protector Cap (secure storage for the docked probes) • •
4.2 mm Probe Tip Thickness Extensions (for better t in outlets) • •
19 mm Probe Tip distance when docked • •
Probe Tip Sheath (UK GS38 sheath–keeps the exposed metal to a
<4 mm limit) • •
Slim Probe for Ultra-Compact Form Factor
Display
LEDs
(All Models) Description
y
x
w
v
u
t
s
Voltage level is backlit
zVoltage level is more than ELV
limit (>50 V ac or >120 V dc)
aVoltage is ac / phase in Single
Pole Phase test
d
DC
e
Voltage is positive or negative at
the indicator probe
bBattery is low / Replace battery
fContinuity or diode in forward
operation
gSwitchable load is ON (two buttons
pressed and current ows)
q r 3-phase sequence indication
detected left or right turning
phases with nonindicator probe
(L1) to indicator probe (L2)
1
3
2
4
5
6
gpn76.eps
LCD
(T130VDE/
T150VDE)
Description
Display is in HOLD mode
Voltage measurement (T130VDE/
T150VDE) or resistance
measurement (T150VDE)
Resistance measurement
(T150VDE)
AC Voltage measurement
DC Voltage measurement
Battery is low / Replace battery
How to Hold the Tester
Always hold the product behind the barrier to keep the
display in view. See Figure 4.
 Warning
To prevent possible electric shock, never
touch the metal pins of the probes when
power is applied.
Self‑Test
The Tester has a built-in self test function.
Before and after use, do a self-test:
1. Touch and hold the probe tips together.
f shows and you can hear the beeper. This
makes sure that the test leads have continuity.
2. Make sure that:
batteries are good
b (T90, T110VDE) is NOT on
(T130VDE VDE) does not show in B , T150
the display
3. Continue to hold the probe tips together for more
than three seconds.
4. Open the probe tips again. All LEDs (all but z
and ) must be on and all symbols in the LCD g
(T130VDE, T150VDE) show for one second. This
test makes sure that all other internal circuits and
indicators are good.
5. Measure a known voltage such as a 230 V socket
outlet. This completes the self-test and includes
the >ELV circuit.
If the Tester fails the self-test or voltage test, do not
use. See To Contact Fluke for servicing.
For an inspection of the insulation, cables, and case,
see .Safety Information
4 5


Product specificaties