Bekijk gratis de handleiding van Kyoritsu KEW 2433R (1 pagina’s), behorend tot de categorie Multimeter. Deze gids werd als nuttig beoordeeld door 52 mensen en kreeg gemiddeld 3.5 sterren uit 26.5 reviews. Heb je een vraag over Kyoritsu KEW 2433R of wil je andere gebruikers van dit product iets vragen? Stel een vraag

Pagina 1/1
INSTRUCTIONMANUAL
KEW SNAP
 
2433R
KEW SNAP Series
LEAKAGECURRENTTESTER
KYORITSU ELECTRICAL INSTRUMENTS WORKS, LTD.,
TOKYO, JAPAN
This instrument has been designed and tested according to IEC
Publication
61010
: Safety Requirements for Electronic Measuring
Apparatus. This instruction manual contains warnings and safety
ruleswhichmustbeobservedbytheusertoensuresafeoperation
ofthe instrument and to retain it in safe condition. Therefore, read
through these operating instructions before starting using the
instrument.
#
WARNING
Readthrough andunderstand instructions containedin this
manualbeforestartingusingtheinstrument.
Save and keep the manual handy to enable quick reference
whenevernecessary.
Theinstrumentistobeusedonlyinitsintendedapplications.
Understandandfollowallthesafetyinstructionscontainedin
themanual.
Failuretofollowtheinstructionsmaycauseinjury,instrument
damage and/or damage to equipment under test. Kyoritsu
is by nomeans liable for any damage resultingfrom the
instrumentincontradictiontothiscautionarynote.
The symbol
#
indicated on the instrument means that the user
mustrefertorelatedpartsof themanual forsafeoperationofthe
instrument. Be sure to carefully read the instructions following
each
#
symbolinthismanual.
#
DANGER
isreservedforconditions andactionsthatarelikely
tocauseseriousorfatalinjury.
#
WARNING
is reserved for conditions and actions that can
causeseriousorfatalInjury.
#
CAUTION
isreservedforconditionsandactionsthatcancause
minorinjuryorInstrumentdamage.
Thefollowingsymbolsareusedandmarkedontheinstrumentand
intheinstructionmanual.Pleasecarefullycheckbeforestartingto
usetheinstrument.
#
Refertotheinstructionsinthemanual.
This symbol is marked where the user must refer to the
instruction manual so as not to cause personal injury or
instrumentdamage.
Indicates an instrument with double or reinforced
insulation.
Indicates that this instrument can clamp on bare
conductorswhen measuringavoltage correspondingto
the applicable Measurement category, which is marked
nexttothissymbol.
IndicatesAC(AlternatingCurrent).
#
DANGER
Never make measurement on a circuit having potential of
300
VACorgreater.
Do not attempt to make measurement in the presence of
flammablegasses.Otherwise,theuse of theinstrument may
causesparking,whichleadstoanexplosion.
Thetransformerjaws aremadeof metalandtheir tipsarenot
completelyinsulated.Beespeciallycarefulaboutthepossible
shorting where the equipment under test has exposed metal
parts.
Neverattempttousetheinstrumentifitssurfaceoryourhand
is
wet.
Do not exceed the maximum allowable input of any
measurementrange.
Never open the battery compartment cover when making
measurement.
Neverfry to make measurement if any abnomal conditions.
suchasbrokenTransformerjawsorcaseisnoted.
The instrument is be used only in its intended applications or
conditions. Otherwise, safety functions equipped with the
instrument doesn
t work, and instrument damage orserious
personalinjurymaybecaused.
Keepyourfingersandhandsbehindtheprotectivefingerguard
duringmeasurement.
#
WARNING
Never attempt to make any measurement, if any abnormal
conditionsarenoted,suchasbrokencase,crackedtestleads
andexposedmetalparts.
Do not install substitute parts or make any modification to
the instrument. Return the instrument to Kyoritsu or your
distributorforrepairorre-calibration.
Do not try to replace the batteries if the surface of the
instrumentiswet.
Always switch off the instrument before opening the battery
compartmentcoverforbatteryreplacement.
#
CAUTION
Make sure that the range selector switch is set to an
appropriatepositionbeforemakingmeasurement.
Do not expose the instrument to the directsun, extreme
temperaturesordewfall.
Besuretosettherangeselectorswitchtothe"OFF"position
afteruse. When theinstrument will not bein use for along
periodoftime,placeitinstorageafterremovingthebatteries.
Usea dampcloth and detergentfor cleaningthe instrument.
Donotuseabrasivesorsolvents.
MeasurementCategory:
Toensuresafe operationof measuring instruments,IEC
61010
establishessafetystandardsforvariouselectricalenvironments,
categorized as O to CAT IV, and called measurement
categories.Higher-numberedcategoriescorrespondtoelectrical
environments with greater momentary energy, so a measuring
instrument designed for CAT III environments can endure
greatermomentaryenergythanonedesignedforCATII.
O :Circuits whichare not directly connected to the mains
powersupply.
CATII :Electrical circuits of equipment connected to an AC
electricaloutletbyapowercord.
CATIII :Primary electrical circuits of the equipment connected
directly to the distribution panel, and feeders from the
distributionpaneltooutlets.
CATIV:The circuit from the service drop to the service
entrance, and to the power meter and primary over-
currentprotectiondevice(distributionpanel).
Incoming wire
:
Device which is
not directly
connected to the
mains power supply
Socket
Interior wiring
2. FEATURES
DigitalclamptesterforACleakagemeasurement.
Accuratetrue-RMSreadingofACcurrentwithdistortedwaveform.
Least affected by external magnetic field, providing wide
measuringrangefromverysmalltolargecurrents.
Designed to safety standard IEC
61010
-
2
-
032
: Measurement
categoryCAT
,
300
Vandpollutiondegree
2
.
Teardrop shapedjawsfor ease of usein crowdedcable areas
andothertightplaces.
Dataholdfunctiontoallowforeasyreadingsindimlylitorhard-to-
reachlocations.
Providesfiltering function toremove high frequency generated by
suchequipmentasinverters.
Peakhold function to allow for measurementof current variation
asshortas
10
msec.
Auto-power-offfunctionpreventsunnecessarypowerconsumption
Dynamicrangeof
4200
countsfullscale.
Largeeasy-to-readLCDdisplaywithletterheightof
13
mm.
Operationconfirmingbeeps.
Insulationbarrieratthetipoftransformerjawsforimprovedsafety.
3. SPECIFICATIONS
Measuringrangesandaccuracy(Sinewave)
Range Resolution MeasuringRange Accuracy(Frequencyrange)
40
mA
0
.
01
mA
0
40
.
00
mA
0〜100A
 ±
1
.
0
%rdg
±
5
dgt (
50
/
60
Hz)
 ±
2
.
5
%rdg
±
10
dgt (
20
Hz
1
kHz)
100〜300A
 ±
1
.
0
%rdg
±
5
dgt (
50
/
60
Hz)
 ±
2
.
5
%rdg
±
10
dgt (
40
Hz
1
kHz)
300〜400A
 ±
2
.
0
%rdg (
50
/
60
Hz)
 ±
5
.
0
%rdg (
40
Hz
1
kHz)
400
mA
0
.
1
mA
0
400
.
0
mA
400
A
0
.
1
A
0
400
.
0
A
CF(Crestfactor)
3
(
45
65
Hz,lessthan
600
APeak)
    ※
100
400
A
sinewave
2
%rdg
Countsequaltoorlessthan
3
countsarecorrectedtozero
Accuracy-insuredFrequencyrangeof
50
/
60
Hzmodeis
50
/
60
Hz.
The max indication at the
40
mA/
400
mA range is
6000
counts.
Minutecurrentmayexistwhilezeroisdisplayedat
400
A/
400
mA
range.Measurementshouldbemadealsoatalowerrange.
Conversion method:
Rmsvaluedetection
Operating System:
Sequentialcomparison
Display:
 LCDwithmax.readingof
4200
(
400
Arange),
6000
(
40
/
400
mArange)
Low battery warning:
"BATT"markappearsonthedisplay
Overrange Indication:
"OL"appearson thedisplay whenupperlimit
ofmeasuringrangeisexceeded
Response Time:
Approx.
2
seconds
Sample Rate:
Approx.
2
.
5
timespersecond
Accuracy-insured
23
℃±
5
,relativehumidity
85
%orless
Temperature and
(withoutcondensation)
Humidity Ranges:
Operating Temperature
0
-
40
,relativehumidity
85
%orless
and Humidity Ranges:
(withoutcondensation)
Storage Temperature
-
20
-
60
,relativehumidity
85
%orless
and Humidity Ranges:
(withoutcondensation)
Operable altitude:
2000
morlessabovesealevel(indooruse)
Power Source:
Two
1
.
5
VR
03
(AAA)batteries
Current Consumption:
Approx.
21
mA
Measurement Time:
Approx.
24
hours
Auto-power-off Function:
Turns power off about
10
minutes after the
lastswitchoperation
Safety Standard:
IEC
61010
-
1
IEC
61010
-
2
-
032
MeasurementCAT
300
V,pollutiondegree
2
EMC directive:
 EN
61326
-
01
,EN
61326
-
2
-
2
Environmetal standard:
EURoHSdirectivecompliant
Overload Protection:
480
AACmax.for
10
seconds
Withstand Voltage:
4240
VACrms (
50
/
60
Hz) for
5
sec. between
metal part of transformer jaws and housing
case(excepttransformerjawcase)
Insulation Resistance:
50
M
Ω
or greater at
1000
V between metal
part of transformer jaws and housing case
(excepttransformerjawcase)
Conductor Size:
Approx.
40
mmindiametermax.
Dimensions:

185
(L)
×
81
(W)
×
32
(D)mm
Weight:
Approx.
270
gincludingbatteries
Accessories:
TwoR
03
(AAA)batteries
CarryingcaseModel
9052
Instructionmanual
Reference
*Effective Value (RMS)
Most alternating currents and
vol tages are expre ssed in
effective values, which are
alsoreferredtoasRMS(Root-
Mean-Square)values.
The eff ective value i s the
square root of the average of
squareofalternatingcurrentor
voltagevalues.Manyclampmetersusingaconventionalrectifying circuit
have"RMS"scalesforACmeasurement.Thescalesare,however,actually
calibratedintermsoftheeffectivevalue ofasinewavethoughtheclamp
meteris respondingto theaverage value.The calibrationis done witha
conversion factor of
1
.
111
for sine wave, which is found by dividing the
effective value by the average value. These instruments are therefore in
erroriftheinputvoltageorcurrenthassomeothershapethansinewave.
*CF (Crest Factor) is found by dividing the peak value by the effective
value.
Examples:
  
Sinewave:CF=
1
.
414
Squarewavewitha
1
:
9
dutyratio:CF=
3
4. INSTRUMENT LAYOUT
●LCD
Low Battery Warning
Frequency Response :
Wide
Frequency Response :
50/60Hz
Data Hold Indication
Peak Hold Indication
Unit
Di s pl ay
Frequency Selector Button
Peak Hold Button
Data Hold Button
Tr a ns fo rm J aw s
Ba r ri er
Tr i gg er
Ra n ge S el ec to r Sw it ch
Ha n d St ra p
Barrier:It is a part providing protection against electrical shock and
ensuringtheminimumrequiredairandcreepagedistances.
5. PREPARATIONS FOR MEASUREMENT
5
-
1
CheckingBatteryVoltage
Setthe RangeSelector Switchto anyposition otherthan theOFF
position.Ifthemarksonthedisplayisclearlyvisiblewithout"BATT"
markshowing,batteryvoltageisOK.Ifthedisplayblanksor"BATT"
is indicated, replace the batteries according to section
8
: Battery
Replacement.
NOTE
When the instrument isleft poweredon,the auto-power-off
functionautomatically shut the power off; Thedisplay blanks
even ifthe Range Selector Switchis set toa position other than
the OFF position in this state. To poweron the instrument, turn
the Range Selector Switch or press the Data Hold Button. If the
display still blanks, the batteries are completelyexhausted.
Replacethebatteries.
5
-
2
CheckingSwitchSetting
MakesurethattheRange SelectorSwitchis settothe appropriate
range.
Also make sure that data hold function is not enabled. If
inappropriate range is selected, desired measurement cannot be
made.
6. OPERATING INSTRUCTIONS
6
-
1
CurrentMeasurement
#
DANGER
Inordertoavoidpossibleshockhazard,nevermakemeasurement
oncircuitshavingapotentialof
300
VACorgreater.
The transformer jaws aremade of metal and theirtips are not
completely insulated.Beespecially careful aboutthe possible
shortingwheretheequipmentundertesthasexposedmetalparts.
Never make measurement withthe batterycompartment cover
removed.
Whenmeasuringcurrentis
300
Aormore(
400
Hzormore),besure
tostopmeasurementwithin
5
minutes.Otherwise,transformer jaws
mayheattocauseafireordeformationofmoldedparts,whichwill
degradeinsulation.
Keepyourfingersandhandsbehindthebarrierduringmeasurement.
#
CAUTION
Take sufficient care to not to apply shock, vibration or
excessiveforcetothejawtips. Otherwise,preciselyadjusted
TransformerJawtipswillbedamaged.
When a foreign substance is stuck in the jaw tips or they
cannot properly engage, the transformer jaws do not fully
close.Insuchacase, donot releasethejawtriggerabruptly
orattemptto closethe transformerjaws byapplying external
force. Make sure that the jaws close by themselves after
removingtheforeignsubstanceormakingthemfreetomove.
The maximum size of a conductor to be tested is
40
mm
indiameter.Accuratemeasurementcannotbemadeona
conductor larger than this, because the transformer jaws
cannotfullyclose.
Whenmeasuringlargecurrent, thetransformerjaws maybuzz.
Thishasnoeffectontheinstrument'sperformanceorsafety.
Sensitivetransformerjaws areusedforLeakageclampmeter.
Because of the characteristics of transformer jaws,which
can be opened and closed, it is impossible to eliminate the
interference of external magnetic field completely. If there
are something,which generatinglarge magnetic field,at a
nearby site, current value can be displayed (
0
cannot
be displayed.) before clamping on the conductor. For such
a case, please use the instrument at a location far from the
thing,whichgeneratingmagneticfield.
Followingarethetypicalthingsgeneratingmagneticfield.
  ・
Conductorfedlargecurrent
  ・
Motor
  ・
Equipmentwhichhasmagnet
  ・
Integratingwattmeter
(
1
)SettheRangeSelectorSwitchtothedesiredposition.Currentto
measureshouldbewithintheselectedmeasuringrange.
(
2
)Normalmeasurement(SeeFig.
1
,
2
):
Pressthejawtriggertoopenthetransformerjawsandclosethem
overoneconductoronly.Measuredcurrentvalueisshownonthe
display.Earthleakagecurrentorsmallcurrentthatflowsthrough
agroundedwirecanalsobemeasuredbythismethod.
(
3
)Measuringoutofbalanceleakagecurrent(SeeFig.
3
):
Clamp onto all conductors except a grounded wire. Measured
currentvalueisshownonthedisplay.
Fig.
1
Loadcurrent
       
Fig.
Earthleakagecurrent
Load
Load
Fig.
3
Measuringoutofbalanceleakagecurrent
Load
Load
3
-phase
3
-wiresystem
(In
4
-wiresystem with neutral,
clampontoall
4
wires)
Single-phase
3
-wiresystem
(in
3
-wiresystem with neutral,
clampontoall
3
wires)
6
-
2
HowtoUseFrequencySelectorButton
Whenhighfrequenciesfromsuchequipmentasinvertersarepresent
inthecircuitundertest,theinstrument measuresAC current ofnot
only
50
Hzor
60
Hzoffundamentalfrequencybutalsoofthese high
frequenciesandharmonics.
To eliminate the effect of such high frequency noise and measure
AC current of
50
Hz or
60
Hz fundamental frequency, a "high-cut"
filtercircuit in incorporated into the instrument which works when
"
50
/
60
Hz" frequency response is selected with the Frequency
SelectorButton. Cut-off frequency of the "high-cut" filter isabout
160
Hzwithattenuationcharacteristicofapprox.-
24
dB/octave.
WhentheFrequencySelectorButtonis pressed,"
50
/
60
Hz"markis
shownontheleftsideofthedisplay.WhentheFrequencySelector
Button is pressed again, fequency response is switchedto WIDE
with"WIDE"markshownonthedisplay.
OutputcharacteristicareshowninFig.
4
.
Fig.
4
KEWSNAP
2433
RFrequencyCharacteristic
-30
-20
-10
0
10
Attenuation dB
k k50 500 10k 50k20k22001002010 100k
100%
50%
20%
10%
5%
Percentage
Typical characteristic "WIDE" frequency response
     fc=8kHz -24dB/cot
Typical characteristic "50/60Hz"
frequency response
     fc=160Hz -24dB/cot
Frequency Hz
Note:
Characteristic of -
24
dB/octiave means that signal magnitude
declinestoaboutonesixteenthofthatattheinitialfrequencywhen
frequency doubles. KEW SNAP
2433
R have the following two
settingsfortheFrequencySelectorButton.
WIDE(
20
Hz-approx.
kHz):
Permits measurement of currents of
fundamental frequencies as well as
currentsofhighfrequenciesgenerated
bysuchequipmentasinverters
50
/
60
Hz(
20
-approx.
160
Hz):Filters out high frequency currents
and measures current of fundamental
frequencyonly
Recently there has been increased use of power through inverters,
switching regulators, etc. When the high frequency noise from such
appliances leaks or flowsinto the ground through capacitors not
filtering completely, the earth leakage breaker may trip even though
thereisno"actual"leakage.Insuchacase,theinstrumentdonotgive
leakagecurrentreadingif"
50
/
60
Hz"frequencyresponseisselected.
Take current readings with the
50
/
60
Hz and WIDE frequency
responses respectively to make effective use of the Frequency
SelectorButton.
6
-
3
PeakCurrentMeasurement
(
1
)SettheRangeSelectorSwitchtothedesiredposition.(Currentto
measureshouldnotexceedtheselectedmeasuringrange.)
(
2
)Select"WIDE"or"
50
/
60
Hz"withtheFrequencySelectorButton.
(
3
)With the transformer jaws clamped onto the conductor under
test,pressthePeakHoldButtonto settheintermenttothepeak
measurementmode.("P"isshownonthedisplay.)
(
4
)The display reads
1
/
2
of thepeak current value. Therefore,an
rms reading is shown when current of a sinusoidal waveform is
measured.
(
5
)Afterpeakmeasurement, pressthe PeakHoldButtontoreturnto
thenormalmeasurementmode.
Note:When leakage current is measured in the peak measurement
mode, the reading may change if the transformer jaws are
openedandclosed.Please readthedisplaywith theconductor
undertest clamped, otherwise, after fixing the display byusing
the data hold function, please remove the instrument from the
conductor to be measured, and read the display. To measure
thepeakcurrentagain,pleasereleasethe datahold, and return
theinstrument to thenormal measurement modeonce with the
PeakHoldButton,thensetitinthepeakmeasurementmode.
Countsequaltoorlessthan
5
countsarecorrectedtozero.
7. OTHER FUNCTIONS
7
-
1
Auto-Power-OffFunction
Thisisafunctionto preventthe instrumentfrom beingleftpowered
onand conserve battery power. The instrument automatically turns
off about
10
minutesafter the last switch or button operation.To
returntothenormalmode,turn theRangeSelectorSwitchtoOFF,
thentothedesiredposition.
DisablingAuto-Power-OffFunction:
Todisabletheauto-power-offfunction,powerontheinstrumentwith
theData Hold Button pressed. About
3
seconds after powering on
theinstrument,"P.OFF"isshownonthedisplay.Toenabletheauto-
power-offfunction,turnontheinstrumentwithout pressingthe Data
HoldButton.
Note: The auto-power-off function is disabled in the peak
measurementmode.
7
-
2
DateHoldFunction
Thisisafunctiontofreezethereadingsonthedisplay.Whenthe
DataHoldButtonispressedonce,thecurrentreadingis held even
thoughcurrent under test varies. "H" mark is shown on the upper
rightcornerofthedisplay.
Toexitthedataholdmode,presstheDataHoldButtonagain.
Note:Whentheauto-power-offfunctionworkswhiletheinstrumentis
inthedataholdmode,dataholdiscancelled.
92
-
1557
H
5
-
22
DISTRIBUTOR
Kyoritsu reserves the rights to change specifications or designs
describedinthismanualwithoutnoticeandwithoutobligations.
8. BATTERY REPLACEMENT
# WARNING
Inorder toavoid possibleshock hazard,always setthe Range
SelectorSwitchtotheOFFpositionbeforetryingtoreplacethe
batteries.
# CAUTION
Donotmixnewandoldbatteries.
Installbatteriesintheorientationasshowninsidethebattery
compartment,observingcorrectpolarity.
Whenthe battery voltage warningmark "BATT" is shownon the
top left corner of the LCD, replace the batteries. Note that the
displayblanksand"BATT"markisnotshownifthe batteries are
completelyexhausted.
(
1
)SettheRangeSelectorSwitchto"OFF."
(
2
)Loosenthebattery-compartment-cover-fixingscrewonthelower
backoftheinstrument.
(
3
)Replace the batteries with two new R
03
(AAA)
1
.
5
V
batteries.
(
4
)Putthebatterycompartmentcoverbackinplaceandtighten
thescrew.
Note:Foruseforalongperiodoftime,usealkalinebatteries(LR
03
).
Screw
Battery
Compartment
Cover
Batteries
DisposetheBatteries
Thismarkingmeanstheyshallbesortedoutandcollectedas
ordainedinDIRECTIVE
2006
/
66
/EC.
This directive is valid only in the EU. When you remove
batteries from this product and dispose them, discard them in
accordance with domestic law concerning disposal. Take a right
actiononwastebatteries,becausethecollectionsystem in the EU
onwastebatteriesareregulated.
WasteElectricalandElectronicEquipment(WEEEE)
Thismarkingmeanstheyshallbesortedoutandcollectedas
ordainedinDIRECTIVE
2006
/
66
/EC.
This directive is valid only in the EU. When you remove
batteries from this product and dispose them,discard them
inaccordancewithdomestic lawconcerningdisposal.Take aright
actiononwastebatteries,becausethecollectionsystem in the EU
onwastebatteriesareregulated.