Bekijk gratis de handleiding van Sanwa DCM401 (2 pagina’s), behorend tot de categorie Meetapparatuur. Deze gids werd als nuttig beoordeeld door 2 mensen en kreeg gemiddeld 4.4 sterren uit 3 reviews. Heb je een vraag over Sanwa DCM401 of wil je andere gebruikers van dit product iets vragen? Stel een vraag

Pagina 1/2
01-2512 2040 6014
− 1 −− 2 −− 3 −− 4 −− 5 −
[1]
SAFETY PRECAUTIONS
Before use, read the following safety
precautions.
This instruction manual explains how to use
your digital clamp meter DCM401. Before use,
when necessary.
Using the product in a manner not specified in
this manual may cause damage to the protection
function of the product.
The instructions given under the headings of
WARNING and
CAUTION must be followed
to prevent accidental and burnelectric shock.
1-1 Explanation of Warning Symbols
manual and attached to the product are as
follows:
: Very important instructions for safe use.
The warning messages are intended to
prevent accidents to operating personnel
such as burn and electric shock.
The caution messages are intended to
prevent incorrect handling which may
damage the product.
: High voltage hazard : Ground
: Resistance : Buzzer
: Light : Capacitor
: Double insulation or reinforced insulation
1-2 Warning Messages for Safe Use
The following instructions are intended to
prevent personal injury such as burn and
electric shock.
Be sure to follow them when using the meter.
1. This is a clamp meter for low-voltage
circuits. Never use it on a power line that
exceeds 1000 Vrms voltage to ground.
2. Never input signals exceeding the
maximum rated input value (see 1-3).
3. Never use the meter near equipment
which generates strong electromagnetic
waves or is charged.
4. Never use the meter if the meter or test
leads are damaged or broken.
5. Never use the meter with the case or
battery lid removed.
6. During measurement, keep your fingers
behind the finger guard of test leads
and the barrier of the meter.
7. During measurement, do not change
function switch of the meter.
8. Before starting measurement, make
sure that the function and range are
properly set.
9. Never use the meter when it is wet or
with wet hands.
10.
Be sure to use the specified type of test leads.
11. Never attempt repair or modification,
except for battery replacement.
12. Always conduct start-up inspection and
check the meter at least once a year.
13. This meter is for indoor use only.
1-3 Maximum Overload Protection Input
protection have been established for the input
terminals of each function.
Function
Signal Input
Terminal
Max. Rated
Input Value
Max. Overload
Protection
ACA,DCA
Clamp type
current
sensor (CT)
DC/AC
1000 Arms
DC/AC
1000 Arms
, ,
Between +
and
terminals
Do not input
voltage
DC/AC
1000 Vrms
ACV, DCV,
LoZ V, EF
DC/AC
1000 Vrms
[2] APPLICATIONS AND FEATURES
2-1 Applications
This is a digital AC/DC clamp meter with DMM functions,
designed for measuring low-voltage circuits rated for CAT.
III 1000 V and CAT. IV 600 V. It is suitable for current and
voltage measurement of low-voltage power lines,
electrical equipment, and power distribution facilities.
2-2 Features
The CT shape makes it easier to clamp cable in crowded.
Complies with IEC61010 CAT. IV and can measure
up to 1000 A AC/DC for use in various sites.
True RMS reading for measurement of even
distorted waveforms.
Low-input impedance voltage measurement
to reduce the effects of ghost voltage.
An electric field (EF) function is incorporated
in the tip of the CT to enable detection of
about AC 90V or more.
Clamp opening diameter of 34 mm at maximum.
Large, easy-to-press HOLD button.
Equipped with LCD backlight convenient for
measurement in dark places. Also incorporates an
LED light to illuminate the measurement target.
Measurement categories (Overvoltage
categories)
CAT. II :
Primary circuit of equipment with a power
cord to be connected to a mains socket.
CAT.III :
Primary circuit of equipment that inputs power
directly from the distributor and the circuit
from the distributor to the mains socket.
CAT.IV :
Circuit from the leading wire to the distributor.
[3] NAMES OF COMPONENT UNITS
3-1 Main unit
3-2 Display
3-3 Test Lead
In case of test pins cover attached:
CAT.IV 600 V, CAT.III 1000 V
In case of test pins cover removed:
CAT.II 1000 V
[4] DESCRIPTION OF FUNCTION
4-1 Power Switch and function switch
(for all functions)
Turn this switch to turn on and off the power
and select a measuring function.
4-2 Data Hold function: HOLD button
(for functions except EF)
When the HOLD button is pressed, the reading
indicated will be held with
on the
display. The indicated reading will not change if
the input signal is changed. When this button is
pressed again, the function will be disabled and
the meter will return to the measurement mode.
Note:
The Data Hold function is released when
the function switch is changed.
4-3 LIGHT function:
Light button
Press the LIGHT button to turn on the display’s
turn them off. The lights will go off automatically
after 30 seconds.
4-4
How to select the measurement function:
SELECT button (ACA,
, ACV)
Press the SELECT button to cycle through the
functions in the order shown below.
 ACA
:
ACAHz → ACA
 
:
 DC/AC/Hz
:
DCVACV → Hz DCV
4-5 Range Hold function:
RANGE HOLD button (ACA, DCA,
, ACV, DCV)
Press the RANGE HOLD button to enter the
Manual mode and set the range (
AUTO
on the
display disappears). The range changes every
time this button is pressed. Select the
appropriate range. To restore the Auto Range
mode, press and hold this button for more than 1
second (
AUTO
is displayed on the display).
4-6
Relative function:
RELATIVE button (ACA, DCA,
 
,
, ACV, DCV)
Press the RELATIVE button; will be
show a relative value which is derived from
the input value when the button is pressed set
at 0 (reference). Press the button again to
cancel this function.
4-7 AUTO POWER OFF function
The meter will go to AUTO POWER OFF function
to save battery life about 10 minutes after the
last function switch or button operation. To wake
up the meter, press any button or re-power on.
turns on while activating the function.
Note:
Even in the AUTO POWER OFF mode, the tiny
power is still consumed. After measurement,
be sure to turn off the power switch.
To disable the function, turn on the power of
meter while holding the SELECT button
pressed.
on the display is turned off when
the AUTO POWER OFF function is disabled.
4-8 Low Battery indication
When the built-in batteries have been
discharged and the voltage has dropped to
below about 3.6 V,
appears on the display.
three batteries with new ones.
[5] MEASURING PROCEDURE
5-1 Start-up Inspection
Always perform the following start-up
inspection to ensure safety.
External appearance: Make sure that there
are no visible abnormalities on the meter that
may have been caused by dropping it or
other mishandling.
Accessories: Make sure that the test leads
and other accessories are free from
breakage, cracks, or any damage.
Make sure that the low battery indication is
not displayed on the display. If
is
displayed, replace the batteries with new
ones. Before first use, install the batteries.
Make sure that neither the meter nor your
hand is wet.
Note:
If nothing is displayed on the display, the
batteries may be drained.
5-2 AC Current / Frequency Measurement:
ACA / Hz
Remove the test leads from the measuring
terminals to avoid electric shock.
FunctionMax. Rated Input Value
ACA1000 A AC
Hz10 kHz
DCA1000 A DC
DIGITAL CLAMP METER
INSTRUCTION MANUAL
DCM401
Note:
Clamp the conductor (cable) to measure in the
middle of the center position mark on the CT.
Clamp only one cable. If several cables are
clamped together or duplex, triplex or
quadruplex cable are clamped, current
cannot be measured accurately.
The meter may malfunction in places where a
strong magnetic field is present.
Do not apply voltage and current at the same time.
5-3 Capacitance Measurement:
Never apply a voltage to the input terminals.
1. Discharge the capacitors before
measurement.
2. This meter uses a measurement method
that applies currents to the measured
capacitor. It is not suitable for measurement
of electrolytic capacitors with high leakage
currents since errors will be too large.
3. It takes longer to measure capacitors with
large capacitance.
FunctionMax. Rated Input Value
60.00 mF
5-4 Resistance Measurement, Continuity
Check:
Never apply a voltage to the input terminals.
FunctionMax. Rated Input Value
60.00 MΩ
600.0 Ω
Dempa Bldg
.
, 4-4 Sotokanda 2-Chome Chiyoda-ku, Tokyo, Japan
- MEMO -
WARNING
WARNING
WARNING
CAUTION
WARNING
Removal test pins cover
Test probe(red)
Test probe(black)
Test pins
Finger
guards
Plugs
LED light
Power /
Function switch
EF detection LED
Barrier
CT trigger
Light button
HOLD button
Display
RANGE HOLD button
SELECT button
RELATIVE button
Input Terminal ( - )
Input Terminal (+)
Clamp type current
sensor (CT)
Electric Field (EF)
sensor position mark
Center position mark
AUTO RANGE
AC
Continuity
check
Polarity
(-)
Decimal
point
Digit
DC
Relative value
Low-impedance
UNIT
DATA HOLD
Low battery
indication
AUTO POWER OFF