Gima 32569 Handleiding
Gima
Niet gecategoriseerd
32569
Bekijk gratis de handleiding van Gima 32569 (4 pagina’s), behorend tot de categorie Niet gecategoriseerd. Deze gids werd als nuttig beoordeeld door 6 mensen en kreeg gemiddeld 4.7 sterren uit 3.5 reviews. Heb je een vraag over Gima 32569 of wil je andere gebruikers van dit product iets vragen? Stel een vraag
Pagina 1/4

M32506-M-Rev.5-02.21
PROFESSIONAL MEDICAL PRODUCTS
STETOSCOPI GIMA - GIMA STETHOSCOPES
STÉTHOSCOPES BIAURICULAIRES GIMA
ESTETOSCOPIOS GIMA - ESTETOSCÓPIO GIMA
GIMA STETHOSKOPE - GIMA SZTETOSZKÓP
STETOSKOPY GIMA - GIMA STETOSKOP
ΣΤΗΘΟΣΚΟΠΙΑ GIMA - GIMA
È necessario segnalare qualsiasi incidente grave vericatosi in relazione al dispositivo
medico da noi fornito al fabbricante e all’autorità competente dello Stato membro in cui
si p1-ha sede.
All serious accidents concerning the medical device supplied by us must be reported to
the manufacturer and competent authority of the member state where your registered of-
ce is located.
Il est nécessaire de signaler tout accident grave survenu et lié au dispositif médical que
nous avons livré au fabricant et à l’autorité compétente de l’état membre où on a le siège
social.
Es necesario informar al fabricante y a la autoridad competente del Estado miembro en el
que se encuentra la sede sobre cualquier incidente grave que haya ocurrido en relación
con el producto sanitario que le hemos suministrado.
É necessário noticar ao fabricante e às autoridades competentes do Estado-membro
onde -ele está sediado qualquer acidente grave vericado em relação ao dispositivo méd
ico fornecido por nós.
Jeder schwere Unfall im Zusammenhang mit dem von uns gelieferten medizinischen Gerät
muss unbedingt dem Hersteller und der zuständigen Behörde des Mitgliedsstaats, in dem
das Gerät verwendet wird, gemeldet werden.
A gyártónak, illetve a székhely szerinti tagállam illetékes hatóságának jelezni kell bármilyen
olyan súlyos balesetet, amely az általunk szállított orvostechnikai eszközzel kapcsolatban
történt.
Należy poinformować producenta i kompetentne władze danego Kraju członkowskiego o
każdym poważnym wypadku związanym z wyrobem medycznym naszej produkcji.
Det är nödvändigt att meddela tillverkaren och de behöriga myndigheterna i den berörda
medlemsstaten, om alla allvarliga olyckor som inträffat i samband med den medicintekni-
ska utrustning som levererats av oss.
Σε περίπτωση που διαπιστώσετε οποιοδήποτε σοβαρό περιστατικό σε σχέση με την ιατρική
συσκευή που σας παρέχουμε θα πρέπει να το αναφέρετε στον κατασκευαστή και στην
αρμόδια αρχή του κράτους μέλους στο οποίο βρίσκεστε.
32506 - 32508 - 32509 - 32511 - 32513 - 32515 - 32526
32527 - 32528 - 32542 - 32569 - 32570 - 32571 - 32572
32574 - 32575 - 32576 - 32577 - 49500 - 49501 - 49502
49503 - 49510 - 49511 - 49512 - 49513
Gima S.p.A.
Via Marconi, 1 - 20060 Gessate (MI) Italy
gima@gimaitaly.com - export@gimaitaly.com
www.gimaitaly.com
Made in Taiwan

4ENGLISH
These medical devices are intended for auscultation.
Bow adjustment
Position the eartubes so that
they are inclined at about 15°
and the eartips are in line with
the nose bridge, facing forward
with respect to you (see illus-
tration).
In this way the eartubes will be
aligned with the ear canal and
you will obtain a clear, sharp
sound.
Holding the eartubes in front
of you with one tube in each
hand, open them, separating
the tubes and inserting the
eartips in your ears.
Adjustment of the bow tension
If you feel uncomfortable or performance is poor, adjust the ten-
sion of the eartubes. Press the binaural eartubes against each
other gently to increase the tension, separate them to reduce it.
• Attention: an excessive tension could weaken the
bow.
• For best performance, it is important to use eartips
of a suitable size for your ears. This is particularly
important if you’re using soft conformable eartips.
Replacement of the diaphragm
Although the diaphragm is solid and planned to last long, it
could be necessary to replace it.
In this case proceed as follows:
• Remove the backing ring, paying attention not to damage it
(for screw models, counter clockwise rotate the ring), then pull
the diaphragm out.
• Fit the new diaphragm and reposition the backing ring making
sure that it is well xed (for screw models clockwise rotate the
ring).

5ENGLISH
Selection of frequency / cushion
for double-head models only
Rotating the chestpiece by 180°, it is possible to change the lis-
tening mode by opening (or indexing) the side you wish to use,
either the diaphragm or open bell: select the diaphragm to listen
to high frequencies, or the open bell to listen to low frequencies.
“If the diaphragm is open, the bell will be closed, blocking the
sound coming from the bell and vice versa. To determine which
of the two sides is active, position the eartips in your ears and
exert slight pressure on the diaphragm or on the hole of the bell.
Cleaning and maintenance
Regular and periodic cleaning will ensure best performance
over time:
• Keep away from extreme heat and cold, solvents and
oils.
• Do not submerge the product in liquids and do not sub-
ject it to sterilisation processes.
• Disinfect the product after each use, using a 70% iso-
propyl alcohol solution.
• Avoid damage to the product caused by sharp objects.
• Regularly unscrew the eartips to keep them clean.
Clean the eartips (and various parts of the product, if
necessary) with alcohol or soapy water and dry them
completely before reassembling them. Make sure the
eartips have been securely retted to the ends of the
eartubes.
• Do not use the product disassembled (e.g. without eart-
ips) and/or in the event of visible damage
GIMA WARRANTY TERMS
The Gima 12-month standard B2B warranty applies.
Product specificaties
Merk: | Gima |
Categorie: | Niet gecategoriseerd |
Model: | 32569 |
Kleur van het product: | Zwart |
Soort: | Akoestische stethoscoop |
Type verpakking: | Doos |
Lengte: | 630 mm |
Membraan diameter: | 45 mm |
Niet-koelrand: | Ja |
Membraandiameter: | 30 mm |
Heb je hulp nodig?
Als je hulp nodig hebt met Gima 32569 stel dan hieronder een vraag en andere gebruikers zullen je antwoorden
Handleiding Niet gecategoriseerd Gima
11 Mei 2025
11 Mei 2025
11 Mei 2025
11 Mei 2025
11 Mei 2025
11 Mei 2025
11 Mei 2025
11 Mei 2025
11 Mei 2025
11 Mei 2025
Handleiding Niet gecategoriseerd
- ProfiCook
- Robbe
- Winia
- CTOUCH
- Gurari
- Vileda
- Finder
- LEDs-ON
- Proficare
- Baby Lock
- Pryme
- Parisot
- National Geographic
- Dynamic
- Advanced Network Devices
Nieuwste handleidingen voor Niet gecategoriseerd
19 Juni 2025
19 Juni 2025
19 Juni 2025
19 Juni 2025
19 Juni 2025
19 Juni 2025
19 Juni 2025
19 Juni 2025
19 Juni 2025
19 Juni 2025