Bekijk gratis de handleiding van Extech 42270 (52 pagina’s), behorend tot de categorie Meetapparatuur. Deze gids werd als nuttig beoordeeld door 64 mensen en kreeg gemiddeld 4.1 sterren uit 9 reviews. Heb je een vraag over Extech 42270 of wil je andere gebruikers van dit product iets vragen? Stel een vraag

Pagina 1/52
User's Guide
Model 42270 Temperature &HumidityDatalogger
Model 42260 Temperature Datalogger
Introduction
Congratulationson your purchase ofthe ExtechDataloggingModule.The following
models are covered in thismanual.
Model 42270: Temperature andHumidity Datalogging Module
Model 42260: Temperature-only Datalogging Module
The Dataloggers can be used to monitor the temperature of greenhouses,
warehouses, food transports, aircraft cabins, refrigerated trucks, containers,
museums, and HVAC equipment. The Model 42270records Relative Humidity
readingsin addition totemperature readings.Instructionson how toprogramthe
dataloggersare not includedthismanual.Programming anddataretrieval
instructionsare includedin the 42275, 42265, 42276 and 42266 kits.
Description
1.Mounting hole
2.Status LEDs (Record andAlarm)
3.LCD display
Displays
LCD Display
REC:Displayedwhile logging readings.
HI and LOW:Displayedwhen the High or Low Alarm limit is exceeded.
RH%:Relative Humidity (42270 only)
C or F:Temperatureunits.
COMM:Appearswhen Datalogger is communicating with a PC.
Recording Status Indication
Both the RED status LED and the display indicator ‘REC’will ash every 5 seconds if
the sampling rate is set to 5 seconds or higher. Ifthe sampling rate is set lower than 5
seconds, the indicators will ashevery 1, 2, 3, or 4seconds asprogrammed.
ALARM Status Indication
Both the YELLOW status LED and the display indicator ‘ALM’will ash when the
recorded value ishigher thanthe userprogrammed HighAlarm value or lower than the
Low Alarm value. The Alarm status LED and the ‘ALM’ displayindicator flash every 5
seconds if the sampling rate is programmedfor 5 secondsor higher. If the sampling
rate is set lower than 5 seconds, the indicatorswill ash every 1, 2, 3, or 4 seconds as
programmed.
Low Battery Indication
When the lithium battery voltage nears the critical operating level, the LCD displays
‘LO’. To replace the battery, refer to the Battery Replacement section of thismanual.
Specications
Front panelstatusLEDs Two (2):RECORD andALARM
Temperature range-40 to 85°C (-40 to 185°F)
Temperature resolution0.1° up to 99.9°; 1° from 100°to 185°
Temperatureaccuracy±0.6°C (1.2°F)from -20 to 50°C (-4 to 122°F)
±1.2°C (2.4°F) all other ranges
Humidity range &accuracy0.0 to 99.9% Relative Humidity; Accuracy: ± 3%
Datalogger Memory16,000 temperature readings (8,000 temperature and
8,000 humidity readings on the 42270)
Sampling rateProgrammablefrom 1 to 7200 seconds (2 hours)
Power supplyOne 3.6V ½ ‘AA’ lithium battery
Battery Life1 year: 5 second sample ratein sleep mode
3.8 months: 24 hour sample rate in non-sleep mode
Low battery IndicatorDisplay shows ‘LO’
Dimensions76.2 x 57.2 x22.3mm (3 x 2.3 x 0.88")
The datalogger display will not switch on
until activated by software or the printer
programmer.
THESE DATALOGGERS CANNOTBEUSEDWITHOUTTHE
SOFTWARE FROM KIT 42275 (42265) OR THE PRINTER
PROGRAMMER 42276 (42266).
Battery Replacement
When the low battery icon (‘LO’) appears on the LCD display, replace the Lithium CR2
battery.
1.Open Datalogger housing by rst removing the four (4) rear Phillips screws.
2.Carefullypry off therear of theDatalogger.
3.The cylindrical lithium battery is locatedon the bottomof the PCB board.
4.Replace the battery, observing polarity.
5.Assemble the Datalogger housing.
You, as the end user, are legally bound (Battery ordinance)to return all
used batteries and accumulators; disposal in the household garbage
is prohibited!
You can hand over your used batteries / accumulators at collection
points in your community orwherever batteries / accumulators are sold!
Disposal: Follow the valid legal stipulations in respect of the disposal of
the device at the end of itslifecycle
Copyright2008 Extech Instruments Corporation©
All rightsreserved including the right of reproduction in whole or in part in any form.
42270_42260-EU-ENV2.2 7/09

Beoordeel deze handleiding

4.1/5 (9 Beoordelingen)

Product specificaties

Merk: Extech
Categorie: Meetapparatuur
Model: 42270

Heb je hulp nodig?

Als je hulp nodig hebt met Extech 42270 stel dan hieronder een vraag en andere gebruikers zullen je antwoorden