Vimar 14503.AB Handleiding


Lees hieronder de 📖 handleiding in het Nederlandse voor Vimar 14503.AB (2 pagina's) in de categorie Niet gecategoriseerd. Deze handleiding was nuttig voor 15 personen en werd door 2 gebruikers gemiddeld met 4.5 sterren beoordeeld

Pagina 1/2
Viale Vicenza, 14
36063 Marostica VI - Italy
www.vimar.com
49400660A0 10 2203
PLANA
14501.AB - 14503.AB - 14504.AB
14501.AB
Call button with red reassurance light - 2 modules.
CHARACTERISTICS.
• Supply voltage: 24 V d.c. ± 20% SELV
• Absorption: 30 mA max
• Recommended installation height: from 0.7 m to 1.5 m
OPERATION.
The device performs the function of calling nursing personnel and patient reassurance
by means of a specic pilot lamp.
CONNECTIONS.
It is connected by traditional wiring to the display module or to the communication
terminal for the room which acts as an interface with the bus line and the landing light
(see the diagram “Connections”).
14503.AB
Cord-operated call button with red reassurance light - 2 modules.
CHARACTERISTICS.
• Supply voltage: 24 V d.c. ± 20% SELV
• Absorption: 30 mA max
OPERATION.
The device is equipped with 1.5 m of cord together with a red disc and it is usually
installed in bathrooms and showers .at least 20 cm above the shower
CONNECTIONS.
It is connected by traditional wiring to the display module or to the communication
terminal for the room which acts as an interface with the bus line and the landing light
(see the diagram “Connections”).
14504.AB
Cancel button with acoustic signal for receiving other calls and green
indicator light - 2 modules.
CHARACTERISTICS.
• Supply voltage: 24 V d.c. ± 20% SELV
• Absorption: 125 mA max
• Recommended installation height: from 0.7 m to 1.5 m
OPERATION.
The device performs the function of cancelling the call and signalling the presence
of personal to the system; it is equipped with a buzzer for the acoustic signal if calls
come from other rooms when the health personnel are not in the control room (calls
are forwarded to the room where the personnel are providing assistance).
CONNECTIONS.
• It is connected by traditional wiring to the display module or to the communication
terminal for the room which acts as an interface with the bus line and the landing
light (see the diagram “Connections”).
• Installare l’apparecchio in una posizione ben accessibile.
The antibacterial treatment ensures full hygiene thanks to the action of the
silver ions (AG+), which prevent the formation and spread of germs, bacteria,
viruses and fungi. To maintain the hygiene and effectiveness of its antibacterial
action, clean the product regularly.
INSTALLATION RULES.
Installation must be carried out by qualied persons in compliance with the current
regulations regarding the installation of electrical equipment in the country where
the products are installed.
STANDARD COMPLIANCE.
EMC directive. EN 60950-1, EN 61000-6-3, EN 61000-6-1 standard.
REACH (EU) Regulation no. 1907/2006 – Art.33. The product may contain traces of lead.
BUZ
G
W
R
G2
BED3L
BED3C
BED2L
BED2C
BED1L
BED1C
OUT24 V d.c. SELV –
WCCL
WCC
WCRL
WCR
NC NC
+
b/–
K2
K1
a/+
b/–
K2
K1
NC
IN+
OUT+
OUT+
OUT+
OUT+
L4 L3 L2 L1 LC
L2 TX L1TS L2 TX L1TS2 B A T 2 B A T
Communication
terminal
02080.AB
or display
module
02081.AB
14342.AB
Bed 2
14342.AB
Bed 3
14501.AB
Bed 1
14503.AB
Bathroom ceiling
pull-cord
14504.AB
Bathroom reset
Lamp
02084
a/+ b/-, : SELV alternating/direct current power supply
K1 K2, : Relay coils for switching on lights with tail call lead
VDE CONNECTIONS.
Viale Vicenza, 14
36063 Marostica VI - Italy
www.vimar.com
49400660A0 10 2203
BUZ
B2LED
BED2C
B1LED
BED1C
WCLED
WCC
WCRLED
WCR
GLAMP
WLAMP
RLAMP
+ OUT
24 V d.c. SELV –
14501.AB
Bed 1
14503.AB
Bathroom ceiling pull-cord
14504.AB
Bathroom reset
14501.AB
Bed 2
NO NO NO NO
Communication
terminal
02080
or display
module
02081
Lamp
L4 L3 L2 L1 LC
02084
14387.R
a/+
b/–
K2
K1
14501 ( 14502)o
Bed 2
14502 ( 14501)o
Bed 1
14503
Bathroom ceiling pull-cord
14504
Bathroom reset
14501 ( 14502)o
Bed 3
14387.B 14387.V
BUZ
G
W
R
G2
BED3L
BED3C
BED2L
BED2C
BED1L
BED1C
OUT24 V d.c. SELV –
WCCL
WCC
WCRL
WCR
+
Communication
terminal
02080.AB
or display
module
02081.AB
Room call Bathroom call Presence
BACKWARD COMPATIBLE CONNECTIONS 1 (OBSOLETE DISPLAY MODULE AND NEW CALL DEVICES AND LAMPS).
BACKWARD COMPATIBLE CONNECTIONS 2 (NEW DISPLAY MODULE AND OBSOLETE CALL DEVICES AND LAMPS).
K1 K2, : Relay coils for switching on lights with tail call
lead (02087 and 02088)
a/+ b/, -: SELV alternating/direct current power supply
02087: K1 - terminals (1-3)
02088: - Left light button: K1 terminals (1-3)
- Right light button: K2 terminals (2-3) The unshielded CAT3 telephone cable can be used for connecting the buttons and lights.
WEEE - User information
The crossed bin symbol on the appliance or on its packaging indicates that the product at the end of its life must be collected separately from other waste. The user must therefore hand the equipment at the end of its life cycle over to the
appropriate municipal centres for the differentiated collection of electrical and electronic waste. As an alternative to independent management, you can deliver the equipment you want to dispose of free of charge to the distributor when
purchasing a new appliance of an equivalent type. You can also deliver electronic products to be disposed of that are smaller than 25 cm for free, with no obligation to purchase, to electronics distributors with a sales area of at least 400
m2. Proper sorted waste collection for subsequent recycling, processing and environmentally conscious disposal of the old equipment helps to prevent any possible negative impact on the environment and human health while promoting
the practice of reusing and/or recycling materials used in manufacture.