Vemer VE 2608 Handleiding
Bekijk gratis de handleiding van Vemer VE 2608 (2 pagina’s), behorend tot de categorie Lamp. Deze gids werd als nuttig beoordeeld door 8 mensen en kreeg gemiddeld 4.8 sterren uit 8 reviews. Heb je een vraag over Vemer VE 2608 of wil je andere gebruikers van dit product iets vragen? Stel een vraag
Pagina 1/2

SAFETY WARNINGS
1)Do not connect the tester to a live voltage point when the switch is set to
Resistance / Diodes / Continuity !
2)Do not take any measurements if any part of the tester or rods is damaged
3)Observe the temperature and humidity limits
4)Do not touch the rods or jacks during measurement operations
5)Do not take measurements in high power circuits with voltage higher than
750 V AC and 1000 V DC
6)Do not power the instrument in an explosive atmosphere
Code Model
VP128300 VE2608
TECHNICAL SPECIFICATIONS
•LCD display: 33/4 digits, maximum reading 4000
•Bargraph: 42 segments, 20 measurements/s
•AC measurements in true effective value (TRMS)
•Polarity: automatic with negative polarity indication
•Automatic zero setting
•Flat battery indication: “ ”
•Measurements by time unit: 2,5 per second (rated)
•Power supply: 1 x 9 Vtransistor battery
•Autonomy: 100 h approx. (alkaline batteries)
•Operating temperature: 0 °C ÷ + 50 °C (0-70% RH)
•Storage temperature: -20 °C ÷ +60 °C (0-80% RH - without batteries)
•Maximum conductor dimensions: Ø max 57 mm (conductor); 70x18 mm (bar)
•Dimensions: 276,5x90,5x47 mm
•Weight: 500 g (batteries included)
•Protection: IP40 (IEC 529 standard)
•Accessories: 2 rods, internal battery, carrying case and user's manual
LEGEND:
A) Dimensions
B)Instrument description
INSTRUMENT DESCRIPTION
Display
Range selector
Negative jack (-)
Insert the black rod
1Positive jack (+) - Voltage measurement
Insert the red rod
2Positive jack - All the other measurements(+)
Insert the red rod
Clamp opening device
Current measurement point
HOLD key
PEAK key
MAX/MIN key
햻“ZERO” key, DC only
햽Bargraph
햾RANGE key
햿Battery compartment
+ -
User's Manual
2000 A AC/DC DIGITAL AMMETRIC CLASMP - BARGRAPH
Read all the instructions carefully
USER'S GUIDE (see panel B)
DIODE CONTROL
1) Cut off power to the circuit to be tested
2) Insert the measurement rods in the tester jacks [black on , red on 2
]
3) To select the diode test, set selector to position “ ”
4) Apply the rods on the diode (red rod with the anode and black with the cathode): if the
diode is operating, a value in the region of 0.6 V should appear on the display
5) “OL”Inverting the rods: if the diode is operating, will appear on the display
6) “0”If the diode is in short circuit, will appear in both tests
CAPACITY
1) Cut off power to the circuit
2) Insert the measurement rods in the tester jacks [black on , red on 2
]
3) To select the capacity test, set selector to position “ ”
4) Connect the two points at which the capacity is to be measured
5) Read the value shown on the diaplay
>1000 MΩ
11 MΩ
10 MΩ
10 MΩ
10 MΩ
Overload protection = 1000 V DC / 750 V AC
INPUT IMPADANCE
DC VOLTS
RANGE
400 mV
4 V
40 V
400 V
1000 V
RESOLUTION
100 Vµ
1 mV
10 mV
100 mV
1 V
TOLERANCE
±(0,5% rdg +1d)
±(0,5% rdg +1d)
±(0,5% rdg +1d)
±(0,5% rdg +1d)
±(0,5% rdg +1d)
(*) Input signal > 40 mV and frequency 50 Hz - 100 Hz
Input impedance same as DCV less than 100pF
Crest factor<=3
Overload protection = 1000 V DC / 750 V AC
AC VOLT 50 Hz - 500 Hz (TRMS)
RANGE
(*) 400 mV
4 V
40 V
400 V
750 V
RESOLUTION
100 Vµ
1 mV
10 mV
100 mV
1 V
TOLERANCE
±(1,5% rdg +4d)
±(1,5% rdg +4d)
±(1,5% rdg +4d)
±(1,5% rdg +4d)
±(1,5% rdg +4d)
ELECTRIC SPECIFICATIONS
(Accuracy calculated to 23 C 5 C < 75% RH)° ±°
Overload protection = 2000 A for 60 s
DC CURRENT
RANGE
400 A
2000 A
RESOLUTION
100 mA
1 A
TOLERANCE
±(1,5% rdg +5d)
400-600 A (1,5% rdg +5d)±
600-800 A (2,5% rdg +5d)±
800-1200 A (3,5% rdg +5d)±
1200-2000 A (5% rdg +5d)±
Overload protection = 2000 A for 60 s
AC CURRENT 50 Hz - 400 Hz (TRMS)
RANGE
400 A
1500 A
RESOLUTION
100 mA
1 A
TOLERANCE
50-60 Hz (1,5% rdg +5d)±
61-400 Hz (3,0% rdg +5d)±
400-600 A
50-60 Hz (1,5% rdg +5d)±
61-400 Hz (3,0% rdg +5d)±
600-1000 A
50-60 Hz (2,0% rdg +5d)±
61-400 Hz (3,5% rdg +5d)±
1000-1500 A
50-400 Hz (5,0% rdg +5d)±
FREQUENCY
RANGE
100 Hz
1 kHz
10 kHz
100 kHz
400 kHz
RESOLUTION
0,01 Hz
0,1 Hz
1 Hz
10 Hz
100 Hz
TOLERANCE
±(0,1% rdg +10d)
±(0,1% rdg +4d)
±(0,1% rdg +4d)
±(0,1% rdg +8d)
±(0,1% rdg +20d)
Min. frequency measurable 1,00 Hz
Overload protection = 500 V DC
For frequences below < 100 Hz and greater than > 100 kHz display may not be stable
RESISTANCE Ω
RESOLUTIONRANGE TOLERANCE
400 Ω
4 kΩ
40 kΩ
400 kΩ
4000 kΩ
40 MΩ
0,1 Ω
1 Ω
10 Ω
100 Ω
1 kΩ
10 KΩ
±(1,2% rdg +4d)
±(1,0% rdg +2d)
±(1,0% rdg +2d)
±(1,0% rdg +2d)
±(1,5% rdg +4d)
±(2,0% rdg +4d)
Overload protection = 500 V
CAPACITANCE
RANGE
4 nF
40 nF
400 nF
4 Fµ
40 Fµ
RESOLUTION
0,001 nF
0,01 nF
0,1 nF
0,001 Fµ
0,01 Fµ
TOLERANCE
±(1,0% rdg +40d) in relative mode
±(1,0% rdg +4d) in relative mode
±(1,0% rdg +4d)
±(1,0% rdg +4d)
± µ(1,0% rdg +4d) for C> 20 F
± µ(5,0% rdg +4d) for C> 20 F
Overload protection = 500 V
CONTINUITY
ACUSTIC SIGNALRANGERESPONSE TIME
400 Ω< 40 circa 100msΩ
Overload protection = 500 V
Overload protection = 500 V
DIODE TEST
TEST CURRENTTOLERANCE
0,6 mA±(1,0% rdg +2d)
OPEN CIRCUIT VOLTS
3,2 V DC
ELECTRIC SPECIFICATIONS
(Accuracy calculated to 23 C 5 C < 75% RH)° ±°
Product specificaties
| Merk: | Vemer |
| Categorie: | Lamp |
| Model: | VE 2608 |
Heb je hulp nodig?
Als je hulp nodig hebt met Vemer VE 2608 stel dan hieronder een vraag en andere gebruikers zullen je antwoorden
Handleiding Lamp Vemer
9 Mei 2023
18 April 2023
14 Maart 2023
1 Maart 2023
Handleiding Lamp
Nieuwste handleidingen voor Lamp
22 Januari 2026
22 Januari 2026
21 Januari 2026
20 Januari 2026
20 Januari 2026
20 Januari 2026
20 Januari 2026
20 Januari 2026
20 Januari 2026
20 Januari 2026