Bekijk gratis de handleiding van PCE Instruments PCE-DC1-ICA (4 pagina’s), behorend tot de categorie Meetapparatuur. Deze gids werd als nuttig beoordeeld door 4 mensen en kreeg gemiddeld 4.4 sterren uit 2.5 reviews. Heb je een vraag over PCE Instruments PCE-DC1-ICA of wil je andere gebruikers van dit product iets vragen? Stel een vraag

Pagina 1/4
Operating Instruction
PCE-DC 1
200A AC/DC Mini Clamp-on Meter
Features
3-2/3 digit (2400 count) LCD display
Built-in non-contact AC voltage detector plus flashlight
Frequency measurement 40Hz to 1kHz
Autoranging with auto power off
0.7 (18mm) Jaw opening
Specifications
z General Specifications
Display 3-2/3 (2400 count) Digit LCD with white LED backlight
Polarity Minus sign ( - ) indicates negative polarity
Current sensor Hall effect sensor type
Overload indication OL displayed on the LCD
DCA zero adjust One touch zero key
Display rate 2 readings/second, nominal
Battery Two 1.5V AAA batteries
Low Battery indication BATT displayed on the LCD
Auto OFF approx. 7 minutes
Operating conditions 32
º
F to 86
º
F (0
º
C to 30
º
C) 90%RH; 86
º
F to 104
º
F (30
º
C to 40
º
C) 75%RH; 104
º
F
to 122
º
F (40
º
C to 50
º
C) 45%RH
Storage conditions - 14
º
F to 140
º
F (-30
º
C to 60
º
C); < 90% Relative Humidity
Altitude Operate at less than 3000 meters
Weight 6.2 oz. (175g) including battery
Dimensions 164 x 65 x 32mm (6.5 x 2.6 x 1.3) (HXWXD)
Jaw opening 18mm (0.7)
Standards IEC61010 Category II 600V,
z Range Specifications
Function Range Resolution Accuracy (of reading)
DC Current 200A 100mA ± (2.8% + 8 digits)
AC Current (50/60Hz) 200A 100mA ±(3.0% + 8 digits)
Frequency
(Auto-Ranging)
40-51Hz 0.01Hz ±(1.2% + 5 digits)
Input Sensitivity: 5A AC RMS min.
51-510Hz 0.1Hz
0.51-1kHz 0.001kHz
Non-Contact AC Voltage 100VAC to 600VAC 50/60Hz
Meter Description
1, Current sense jaw
2. Flashlight button
3. Rotary function switch
4. Data HOLD and Backlight key
5. LCD display
6. Clamp trigger
7. One-touch DCA ZERO key
8. Non-contact AC voltage indicator light
9. Non-contact AC voltage detector probe tip
Safety Information
!
!
!
!!
Caution! Refer to the explanation in this Manual
Double Insulation
This meter has been designed to be safe in use, but the operator must use caution in its operation. The
rules listed below should be carefully followed for safe operation.
1. NEVER apply voltage or current to the meter that exceeds the specified maximum:
2. USE EXTREME CAUTION when working with voltages greater than 25VAC rms. These voltages are
considerd a shock hazard.
3. NEVER operate the meter unless the back cover and the battery/fuse door are in place and fastened
securely
Input Limits
C
C
AC/DC CLAMP METER
AC/DC CLAMP METER
Function Maximum Input
AC Current 200A
DC Current ±200A
Frequency 1kHz
Operation
AC Current Measurements
1) Set the Function switch to the ACA range.
2) Press the jaw trigger and clamp around,
fully enclosing a single conductor.
Do not allow a gap between the two halves of the jaw.
Refer to the diagram at right for
the correct way to enclose a single conductor.
3) Read the ACA value on the LCD.
DC Current Measurements
1) Set the Function switch to the DCA range.
2) Press the ZERO key to null the meter display.
3) Press the Trigger to open the current sense Jaw.
4) Fully enclose the conductor to be measured.
Do not allow a gap between the two halves of the jaw.
5) Read the DCA value on the LCD.
Non-Contact AC Voltage Measurements
1) Touch the probe tip to the hot conductor or insert into the hot side of the electrical outlet.
2) If AC voltage is present, the detector light will illuminate.
NOTE: The conductors in electrical cord sets are often twisted. For best results, rub the probe tip along
a length of the cord to assure placing the tip in close proximity to the live conductor.
NOTE: The detector is designed with high sensitivity. Static electricity or other sources of energy may
randomly trip the sensor. This is normal operation
Frequency Measurements
1) Set the Function switch to the Hz position.
2) Fully enclose the conductor to be measured. Do not allow a gap between the two halves of the jaw.
3) Read the Frequency value on the LCD.
Flashlight
Press and hold the top button to turn the flashlight on. Release the button to turn the flashlight off.
WARNING: Risk of Electrocution. Before use, always test the Voltage Detector on a
known live circuit to verify proper operation