Bekijk gratis de handleiding van PCE Instruments PCE-CPC 50 (24 pagina’s), behorend tot de categorie Meetapparatuur. Deze gids werd als nuttig beoordeeld door 25 mensen en kreeg gemiddeld 4.5 sterren uit 13 reviews. Heb je een vraag over PCE Instruments PCE-CPC 50 of wil je andere gebruikers van dit product iets vragen? Stel een vraag
Pagina 1/24

User Manual
Cleanroom Particle Counter
PCE-CPC 50

1/ 23
Cleanroom Particle Counter PCE-CPC 50
Applications
Power battery manufacturing
Chip manufacturing
Medical and pharmaceutical
Precision machining
Description
The online the PCE-CPC 50 particle counter adopts principle of optical scattering, which can accurately detect and
calculate suspended the number of particles of different particle sizes in the air per unit output volume. It can the
particle count of 6 channels of 0.5μm, 1.0μm, 2.5μm, at unit 0.3μm, 5.0μm and 10μm the (the same time default is
pcs/m³, can be switched to pcs/L; pcs/28.3L).
Features
4 in 1 integrated particle counter of particulate matter sensor, filter, flow sensor and pump
Output particle number (PCS/m³, PCS/L or PCS/28.3L) in 6 channels including 0.3μm, 0.5μm, 1.0μm, 2.5μm,
5.0μm, 10μm
Built-in high power industrial grade linear laser, accurate identification
Built-in pump and flow sensor for constant stable sampling
Wide working temperature -30 ~70 , stable for different clean room application℃ ℃
With voltage regulator design and EMC compliant, strong anti-static ability
Modbus RS485,MQTT, 4-20mA output for online remote monitoring
Working Principle
When sampled particles pass through light beam (laser), there will be light scattering phenomenon. Scattered light
will be converted into electrical signal (pulse) via photoelectric transformer. The bigger particles will obtain stronger
pulse signal (peak value). Through peak value and pulse value quantity concentration of particles in each size can
be calculate. Thus, real-time measurement data is obtained through measuring quantity and strength of scattered
light.

2/ 23
Specification
Cleanroom Particle Counter PCE-CPC 50 Specification
Operating principle
Light Scattering
Detect particle diameter range
>0.3μm, >0.5μm, >1.0μm, >2.5μm, > 5.0μm, >10μm
Measurement error
<100pcs/l: 30pcs/L±
>100pcs/L: 30% of reading±
Condition: 0C ~ 40C,50+10%RH
(0°C ~ 40°C, 50 ±10%RH. Reference instrument TSI 9306. TSI 9306
count efficiency 50% @ 0.3 µm and 100% for particles >0.45 µm)
Ti
me
to
first reliable reading
≤8s
Sampling interval
1s
Working condition
0 45 ; 0 95%RH (non-condensing)℃~ ℃ ~
Storage condition
-20~60 , 0~95%RH (non-condensing)℃
Working current
≤1A
Communication
RS485 (standard)
RJ45 (standard)
4~20mA (standard)
Lifetime
≥3 years
Sampling flow rate
2.83L/min
Standard
Comply with JJF 1190-2008
Product specificaties
Merk: | PCE Instruments |
Categorie: | Meetapparatuur |
Model: | PCE-CPC 50 |
Heb je hulp nodig?
Als je hulp nodig hebt met PCE Instruments PCE-CPC 50 stel dan hieronder een vraag en andere gebruikers zullen je antwoorden
Handleiding Meetapparatuur PCE Instruments
5 Juli 2025
4 Juli 2025
17 Juni 2025
16 Juni 2025
16 Juni 2025
15 Juni 2025
15 Juni 2025
25 Mei 2025
22 Mei 2025
22 Mei 2025
Handleiding Meetapparatuur
- Gude
- Milwaukee
- Hikmicro
- Voltcraft
- Flex
- Stamos
- Bosch
- Fixpoint
- Livington
- VDO
- Kübler
- Jonard Tools
- EBERLE
- ORNO
- Silverline
Nieuwste handleidingen voor Meetapparatuur
29 Juli 2025
29 Juli 2025
29 Juli 2025
29 Juli 2025
29 Juli 2025
29 Juli 2025
28 Juli 2025
28 Juli 2025
28 Juli 2025
28 Juli 2025