Bekijk gratis de handleiding van Extech CO220 (10 pagina’s), behorend tot de categorie Niet gecategoriseerd. Deze gids werd als nuttig beoordeeld door 27 mensen en kreeg gemiddeld 5.0 sterren uit 14 reviews. Heb je een vraag over Extech CO220 of wil je andere gebruikers van dit product iets vragen? Stel een vraag

Pagina 1/10
USERGUIDE
CO2   Monitor and Datalogger
ModelCO220
2 CO220-en-GB_V1.3 07/13
Introduction
Congratulationson ofyourpurchase the CO220 CarbonModel  DioxideMeter.Thismeter
measuresCO2
concentration,airtemperature,andrelative includeshumidity.TheCO200  a 99
Memory timeDataloggerforstoringand andretrieving date readings with   stamping.
The alarms foraudible makethisisanidealinstrument indoorairquality (IAQ) diagnostics.
Carbondioxide(CO2) ais gaseouscomponentoftheearth’satmosphere.Theconcentrationof
CO2   in natural ambient air is approximately 0.04% or 400ppm.
This tested calibratedmeterisshippedfully and and,withproperuse,willprovideyearsof
reliableservice. forPleasevisit (www.extech.com)our website tocheck thelatestversionofthis
UserGuide,ProductUpdates, Customerand Support.
FEATURES
Extra   Large LCD simultaneously displays CO2   Level, Air Temperature, & Relative Humidity
DisplayedCalendarClock.
Six           (6) icons indicate indoor air quality levels (350 / 450 / 700 / 1000 / 1250 / 5000ppm)
Stable  NDIR sensor for CO2 detection
Automatic   Baseline Calibration (ABC) function
Audible high CO2     concentration alarm with selectable high alarm threshold
Dew   Point and Wet Bulb temperature measurements
Descriptions
METER
1. CO2 reading
2. RelativeHumidityReading
3. AirTemperatureReading
4. Time/Date Display
5. SETbutton
6. ESC – MEMbutton
7. SELECT – RECALLbutton
8. C/F‐%RHbutton
9. CO2 button
10. ACadaptorjack
NOTE:CO2,TemperatureandRHsensorsonrearofinstrument
3 CO220-en-GB_V1.3 07/13
DISPLAYSYMBOLDESCRIPTIONS
ppmCO2   concentration unit of measure
  350ppm to 450ppm
  450ppm to 700ppm
  700ppm to 1000ppm
  1000ppm to 2500 ppm
  2500ppm to 5000ppm
  5000ppm and above
AirTemp Air Temperature
TWA   Time Weighted Average (8 hrs)
STEL   Short Term Exposure Limit (15 minute weighted average)
IPS%    Liters per second per person
cfm/p     Cubic feet per minute per person
WBWet  Bulb temperature
DP   Dew Point temperature
MAX/MIN   Maximum or Minimum reading
PUSHBUTTONDESCRIPTION
SET   Enter setup mode
Saveandconfirmsettings
ESCMEM   Exit setup and programming pages
End  a calibration session
Memory   Datalogger (99 points)
SELECT/RECALL   Select/Recall and Clear Datalogger Memory Readings
oC/oF‐%RH   Temperature/Relative Humidity Modes/Temperature Unit changing
CO2CO2   mode page stepping