Extech 433201A Handleiding
Extech
Meetapparatuur
433201A
Bekijk gratis de handleiding van Extech 433201A (8 pagina’s), behorend tot de categorie Meetapparatuur. Deze gids werd als nuttig beoordeeld door 14 mensen en kreeg gemiddeld 4.3 sterren uit 7.5 reviews. Heb je een vraag over Extech 433201A of wil je andere gebruikers van dit product iets vragen? Stel een vraag
Pagina 1/8

User's Guide
Microprocessor Calibrator/Thermometer
Models 433201A and 433202A

433201A-EN v1.6 07/13
2
Introduction
Congratulations on your purchase of the Extech 433201A (115V) or 433202A (220V)
Microprocessor Thermocouple Calibrator/Thermometer.
In the ‘Measurement’ mode, a millivolt (mV) signal is accepted from a thermocouple and is
displayed on the 4-digit LCD display. A built-in temperature offset feature can be used to
compensate for variations found in standard thermo he offset, measurement couples. By adjusting t
accuracy for a particular thermocouple can be optimized for a particular temperature.
In the ‘Source’ mode, a millivolt signal is output to an external device with the corresponding
temperature (according to standard temperature tables) displayed on the LCD display. Preset
temperature settings (source mode only) according to thermocouple type are stored in five memory
locations for easy recall.
If desired, the meter can be switched to the mV mode where Measurement and Source values
can be displayed in millivolts rather than temperature.
Careful use of this device will provide years of reliable service.

433201A-EN v1.6 07/13
3
Description
1. LCD display – Temperature values can be displayed in
oC/oF or in millivolts. The adjustable
display can be easily moved to the best viewing angle.
2. POWER/MODE button – Press to power meter on/off and to select the desired mode of
operation. With the meter powered on, momentarily press the POWER/MODE button (<2
sec) to toggle between MEASURE and SOURCE mode. In MEASURE mode the meter
displays the temperature or mV value for an attached thermocouple or thermocouple
simulator. In the SOURCE mode, the meter outputs a mV signal.
To power off, press and hold the POWER/MODE button for more than 2 seconds.
3. FINE/COARSE output adjustment buttons – Use the ▼ ▲/ buttons to set the desired
temperature. Press and hold for COARSE adjustment; momentarily press for FINE
adjustment.
4. oC/oF button – Press to select the desired unit of measure
5. TYPE button – Press to select a thermocouple type
6. MEM – In SOURCE MODE, the Memory button provides access to five memory locations
with preset output values depending on which thermocouple type is being used.
Memory
Location
Type
J, K, C, R, S, N
Type T Type E mV
M1 32.0oF
(0.0o
C)
32.0oF
(0.0oC)
32.0oF
(0.0oC)
0mV
M2 212.0o
F
(100.0oC)
212.0o
F
(100.0oC)
212.0o
F
(100.0oC)
10.00mV
M3 932.0
oF
(500.0oC)
392.0oF
(200.0oC)
482.0oF
(250.0oC)
25.00mV
M4 1382.0 o
F
(750.0oC)
572.0oF
(300.0oC),
932.0oF
(500.0oC)
40.00mV
M5 1832.0 o
F
(1000.0oC)
752.0oF
(400.0oC)
1382.0 o
F
(750.0oC)
50.00mV
7. Output/Input mini connector – Thermocouples, thermocouple wire, and calibration cables are
connected here.
8. AC adaptor/recharger jack
9. Battery compartment – Houses the supplied 9V Nickel Metal Hydride rechargeable battery.
An alkaline battery can also be used.
10. Neck Strap Posts- Attach the neck-strap to the two posts on the front corners of the meter.
Product specificaties
Merk: | Extech |
Categorie: | Meetapparatuur |
Model: | 433201A |
Heb je hulp nodig?
Als je hulp nodig hebt met Extech 433201A stel dan hieronder een vraag en andere gebruikers zullen je antwoorden
Handleiding Meetapparatuur Extech
24 Mei 2025
24 Mei 2025
24 Mei 2025
1 Mei 2025
1 Mei 2025
1 Mei 2025
1 Mei 2025
1 Mei 2025
17 Maart 2025
30 Januari 2025
Handleiding Meetapparatuur
- Mitutoyo
- RIDGID
- Ade
- SIIG
- Westfalia
- Ebro
- Yato
- Powerfix
- Jonard Tools
- Draper
- CESYS
- Elgato
- Voltcraft
- Eltako
- Maximum
Nieuwste handleidingen voor Meetapparatuur
16 Juni 2025
16 Juni 2025
16 Juni 2025
15 Juni 2025
15 Juni 2025
15 Juni 2025
15 Juni 2025
15 Juni 2025
15 Juni 2025
15 Juni 2025