Bekijk gratis de handleiding van Extech 380580 (8 pagina’s), behorend tot de categorie Niet gecategoriseerd. Deze gids werd als nuttig beoordeeld door 34 mensen en kreeg gemiddeld 5.0 sterren uit 17.5 reviews. Heb je een vraag over Extech 380580 of wil je andere gebruikers van dit product iets vragen? Stel een vraag

Pagina 1/8
USER Manual
Model 380580
Battery Powered Milliohm Meter
2 - - _v2.7 /22 380580 en US 10
Introduction
Congratulations on you Model 380580 Battery Powered Milliohm Meter. This device r purchase of Extech’s
offers five resistance ranges with resolution as low as 0.1m . The 4-wire Kelvin clip connection ensures
optimum accuracy. Typical applications include transformer, motor coil, and PC Board resistance
measurements. This professional meter, with proper care, will provide years of safe reliable service.
Meter Description
1. Current Terminals
2. Potential Measurement Terminals
3. LCD Display
4. Start/Stop Test Button
5. Range Select/Power Switch
6. LED Error Lights
No Test/Over Temperature
R
P
Voltage Regulation
R
C
Current Regulation
7. Current/Range Table
Leads
Current Leads- Banana plug to alligator clip
- Green C1
- Blue C2
Voltage Potential Leads- Banana plug to alligator clip
- Red P1
- Black P2
Kelvin Clips- Banana Plugs (2) to Kelvin Clip
Red (P1) Green (C1)
Black (P2) and Blue (C2)
Measurement Considerations
1. Do not apply voltage to the meter input terminals. Meter damage may result.
2. Always insure that the circuit to be measured is switched OFF, isolated and completely de-energized
before connecting the test leads.
3. If the Over Temperature LED (NO TEST) indicator is lit, allow the instrument to cool down before
proceeding further.
4. The R
C
led indicates when the test current falls out of regulation. Selecting a higher range may
eliminate the condition.
5. The R
P
led indicates when the voltage on the device under test is too high. Selecting a lower range
may eliminate the condition.
6. If either the R
C
or R
P
led is on, the measurement may be in error.
7. The current terminals are fuse protected.
8. Keep the potential test leads as short as possible. Long leads may introduce noise.
9. When using the four separate alligator clip leads always place the current leads outside the potential
leads.
3 - - _v2.7 /22 380580 en US 10
Preliminary Checks
Current Regulation Check
1. Connect the current leads C
1
and C
2
to the meter.
2. Set the function switch to the 200.0m range. Ω
3. Momentarily press the TEST R
P
button. The meter will
intermittently beep Rand
C
will light
4. Short the current leads C
1
to C
2.
5. The R
C
LED should go off, indicating that the meter is operating
correctly.
6. Momentarily press the TEST R
P
button to stop the test
7. The meter will return to status. NO TEST
Voltage Measurement check
1. With the current test leads C
1
and C
2
shorted, connect and short
the potential (voltage) leads P
1
and P
2
.
2. Set the Function switch to the 200.0m position. The ΩNO TEST
status LED will light.
3. Momentarily press the TEST R
P
button. (the meter will
intermittently beep)
4. The display should indicate 00.0
5. Momentarily press the TEST R
P
button to stop the test.
The status LED will light. NO TEST
6. Remove the shorts from P
1
and P
2
, and C
1
and C
2
and
7. Short the test leads P
1
to C
1
and P
2
to C
2
8. The R
P
LED as well as the status LED should light NO TEST
indicating an over-voltage or over-range
9. Turn the rotary selector switch to OFF
Polarity check
1. Short the test leads P
1
Cto
2
Pand
2
Cto
1
together.
2. Set the Function switch to the 200.0m position. Ω
3. The “ “ negative indicator should appear in the display.-1
Operation check
Use the Kelvin clips for this test
1. Short the all the leads (P
1
, P
2
,C
2
,C
1
) together. The P
1
, P
2
,C
2
,C
1
order is important.
2. Set the Function switch to the 200.0m position. The ΩNO
TEST status LED will light.
3. Momentarily press the TEST R
P
button. (the meter will
intermittently beep
4. The display should indicate near 00.0 ( -0.2m depending +/ )
on the test clip connections and both R
P
and R
C
LEDs should
remain off.
5. Momentarily press the TEST R
P
button to stop the test.
Note: These tests can be performed on any range.


Product specificaties

Merk: Extech
Categorie: Niet gecategoriseerd
Model: 380580

Heb je hulp nodig?

Als je hulp nodig hebt met Extech 380580 stel dan hieronder een vraag en andere gebruikers zullen je antwoorden




Handleiding Niet gecategoriseerd Extech

Extech

Extech CG206 Handleiding

26 Augustus 2025
Extech

Extech TKG100 Handleiding

25 Augustus 2025
Extech

Extech TKG150 Handleiding

25 Augustus 2025
Extech

Extech TKG250 Handleiding

25 Augustus 2025
Extech

Extech 380580 Handleiding

22 Oktober 2024
Extech

Extech HDV540 Handleiding

11 December 2023

Handleiding Niet gecategoriseerd

Nieuwste handleidingen voor Niet gecategoriseerd