Bekijk gratis de handleiding van Daytronic AA14-4F010 (8 pagina’s), behorend tot de categorie Niet gecategoriseerd. Deze gids werd als nuttig beoordeeld door 12 mensen en kreeg gemiddeld 5.0 sterren uit 4 reviews. Heb je een vraag over Daytronic AA14-4F010 of wil je andere gebruikers van dit product iets vragen? Stel een vraag
Pagina 1/8

1GENERALDESCRIPTIONANDSPECIFICATIONS
The AA14-4F010 is a high-isolation, four-channel conditioner, accepting real-world
temperature signals from
Types E, J, K, N, R, S, and T Thermocouples. Based on
NISTpolynomials, linear output is produced over each thermocouple’s stated operat-
ing range, and within the rated accuracy limits—
without the need for additional output
processing.
See Table 1 for the “mV/degree” linear output produced by the AA14-
4F010 for different TCtypes and ranges (°C or °F).
PLEASENOTE: WHILE ALL FOUR CHANNELS OF A GIVEN AA14-4F010 CARD MUST
BE DEDICATED TO THE SAME TC TYPE
, INDIVIDUAL CARDS DEDICATED TO DIFFER-
ENT TC TYPES MAY BE MIXED AS DESIRED WITHIN THE TOTAL DATA ACQUISITION
SYSTEM. ITISALSOREQUIREDTHATTHE AA14-4F010’S
CHANNELNO. 1BEUSED.
This conditioner features
true galvanic isolation with pulse-width modulation,
allowing sensor-to-chassis or sensor-to-sensor common-mode voltages as high as
1500 V (rms) to be accommodated. Internal reference-junction compensation is auto-
matically selected by thermocouple choice. No external cold junction is required
(although the user may supply his own Controlled Ambient Temperature Zone for ref-
erence-junction purposes, if desired).A conventional Daytronic four-channel isother-
mal connector assembly (No. 60323, shown in Figs. 2 and 3) is supplied with each
Model AA14-4F010, with screw terminals for direct connection of TC leads (which can-
not be soldered), and with a
precision thermistor for measurement of the reference-
junction temperature. The same connector may be used with any TC type.
In the event of a broken thermocouple wire or other “open TC” condition, the AA14-
4F010 will automatically report an indeterminate off-scale reading for the TC channel
in question, with positive or negative polarity selectable on a per-channel basis (as
explained in Section 3.a).
Each AA14-4F010 input channel employs an active low-pass filter with a fixed cutoff
frequency of 10 Hz. A nominal ±5-V ANALOGOUTPUT is produced by each active
AA14-4F010 input channel, for purposes of real-time signal monitoring. Each of these
“Auxiliary Outputs” is available on a corresponding mainframe wire-wrap pin. Each
output may be individually set, if desired, to represent the
prefiltered value of the cor-
responding input (see Section 3.b).
“Absolute” calibration lets you quickly set up each TC-based data channel simply by
entering appropriate “type” and “range” information. During operation, appropriate
reference-junction compensation, real-time digital linearization, and engineering-unit
scaling are automatically applied for each type of thermocouple used. A second, two-
point “zero and span” calibration technique is provided, however, for applications
where it is desirable to force multiple TC readings to the same exactly known temper-
ature.*
MODELAA14-4F010
THERMOCOUPLE
CONDITIONERCARD
GENERALDESCRIPTIONANDSPECIFICATIONSAA14-4F010.1
SYSTEM10
ANALOGINPUTCARDS
*In heat balance applications, for example, it is useful to be able to zero all readings simultane-
ously.
1.E.2.AA14-4F010
Product specificaties
| Merk: | Daytronic |
| Categorie: | Niet gecategoriseerd |
| Model: | AA14-4F010 |
Heb je hulp nodig?
Als je hulp nodig hebt met Daytronic AA14-4F010 stel dan hieronder een vraag en andere gebruikers zullen je antwoorden
Handleiding Niet gecategoriseerd Daytronic
25 Januari 2024
25 Januari 2024
25 Januari 2024
25 Januari 2024
25 Januari 2024
25 Januari 2024
25 Januari 2024
25 Januari 2024
25 Januari 2024
25 Januari 2024
Handleiding Niet gecategoriseerd
Nieuwste handleidingen voor Niet gecategoriseerd
26 Augustus 2026
26 Augustus 2026
26 Augustus 2026
26 Augustus 2026
26 Augustus 2026
26 Augustus 2026
26 Augustus 2026
26 Augustus 2026
26 Augustus 2026
26 Augustus 2026