Bekijk gratis de handleiding van Daytronic 10A74-4C (18 pagina’s), behorend tot de categorie Niet gecategoriseerd. Deze gids werd als nuttig beoordeeld door 29 mensen en kreeg gemiddeld 4.2 sterren uit 3 reviews. Heb je een vraag over Daytronic 10A74-4C of wil je andere gebruikers van dit product iets vragen? Stel een vraag
Pagina 1/18

1GENERALDESCRIPTIONANDSPECIFICATIONS
Intended primarily for stress analysis applications, the Model 10A74-4C receives and
conditions up to
four independent inputs from one-arm (1/4-bridge), two-arm (1/2-
bridge), and four-arm (full-bridge)
strain gage configurations attached directly to the
stressed material.
If you are measuring strain by means of a one-arm (1/4-bridge) strain gage configura-
tion, BE SURE TO STUDY SECTION 4 FOR A DISCUSSION OF “LEAD WIRE AND NON-
LINEARITY EFFECTS WITH QUARTER-BRIDGE STRAIN GAGE CONFIGURATION.”
To minimize “time-skew” effects, each 10A74-4C input has a separate TRACK/HOLD
circuit. Upon application of a “HOLD” command, the channel will instantly freeze its
present value in analog memory until application of a “TRACK” command, whereupon
normal tracking of the input signal is resumed. By applying a “HOLD” command
simultaneously to all four channels, the user can be assured of a set of readings
coherent to within a few microseconds.
As explained in Section 3.e, the TRACK/HOLD function is normally controlled through
an external logic signal to the 10A74-4C’s rear I/O connector, when a System 10
Bridge Completion Connector or the optional Model 10VAC is not being used. When
the conditioner is used in a “B-sized” System 10 mainframe (only), the TRACK/HOLD
function may be alternatively controlled through an optional Model 10A1 Control Sig-
nal Card mounted in the same “A-card” rack.* Setup of the 10A1 for use with the
10A74-4C is discussed in Section 3.b.
Unless all four inputs originate from full-bridge transducers, the 10A74-4C requires
external completion of each strain gage input to
full bridge, by means of one of the fol-
lowing:
•aSystem 10 Bridge Completion Connector—one of the following models:
—
10QBC-4-120 for 2- or 3-wire 120-Ωquarter-bridgeconnections
—
10QBC-4-350 for 2- or 3-wire 350-Ωquarter-bridgeconnections
—
10QBC-4-1K for 2- or 3-wire1000-Ωquarter-bridge connections
—
10HBC-4 for 4- or 6-wire half-bridge connections
—
10FBC-4 for 5-wire full-bridge connections
•a Daytronic Model 10CJB-4 Quad Bridge Completion Card(described in Sec-
tion 2.b);
•equivalent circuitry supplied by the user**
*In System 10, the Model 10A1 uses the same system bus as the Model 10A11 Thermocouple
Processor Card; therefore, a “10A” THERMOCOUPLE CONDITIONER such as the Model
10A9-8C or 10A10-4 cannot be used in the same System 10 “A-card” rack as a Model 10A1.
**If you wish to provide your own bridge-completion circuitry, you should contact the factory for
instructions on connections and calibration.
MODEL10A74-4C
QUADSTRAINGAGETRACK-HOLD
CONDITIONERCARD
GENERALDESCRIPTIONANDSPECIFICATIONS10A74-4C.1
1.E.2.10A74-4C
SYSTEM10
ANALOGINPUTCARDS
Product specificaties
| Merk: | Daytronic |
| Categorie: | Niet gecategoriseerd |
| Model: | 10A74-4C |
Heb je hulp nodig?
Als je hulp nodig hebt met Daytronic 10A74-4C stel dan hieronder een vraag en andere gebruikers zullen je antwoorden
Handleiding Niet gecategoriseerd Daytronic
25 Januari 2024
25 Januari 2024
25 Januari 2024
25 Januari 2024
25 Januari 2024
25 Januari 2024
25 Januari 2024
25 Januari 2024
25 Januari 2024
25 Januari 2024
Handleiding Niet gecategoriseerd
Nieuwste handleidingen voor Niet gecategoriseerd
26 Augustus 2026
26 Augustus 2026
26 Augustus 2026
26 Augustus 2026
26 Augustus 2026
26 Augustus 2026
26 Augustus 2026
26 Augustus 2026
26 Augustus 2026
26 Augustus 2026