Bekijk gratis de handleiding van Daytronic 10A68-2 (4 pagina’s), behorend tot de categorie Niet gecategoriseerd. Deze gids werd als nuttig beoordeeld door 20 mensen en kreeg gemiddeld 4.4 sterren uit 4 reviews. Heb je een vraag over Daytronic 10A68-2 of wil je andere gebruikers van dit product iets vragen? Stel een vraag
Pagina 1/4

1GENERALDESCRIPTIONANDSPECIFICATIONS
Using the special screw-terminal connector board described in Section 2, the Model
10A68-2 accurately measures the true RMS amplitude of one or two independent
external analog AC signals, mixed as desired, from a wide range of voltage or cur-
rent sources
—including engineering parameters conditioned by other Daytronic
cards.
Since the 10A68-2 does not provide excitation, the analog source must supply its own
power supply
, if required (as is the case with many accelerometers, when used to
measure the noise level or vibration factor of a dynamic process). The card is particu-
larly useful in evaluating the
efficiency of AC electrical power systems.
ADDITIONAL10A68-2 SPECIFICATIONS
Transducer Types
: AC signal sources (voltage or current)
Input Voltage and Current Ranges: See Table 1, below, for allowable full-scale input
ranges (given for each range are the corresponding
normal-mode input impedance
(z
i
), maximum normal-mode input voltage or current with no instrument damage
(V
max
or I
max
), and bandwidthflat to 0.1%, 1%, and 3 dB, respectively); automatically
selected—on an individual channel basis—when the channel is configured; for the
System 10 channel “type” codes assigned to 10A68-2 data channels, see Table 2,
below
Amplifier (per channel): Floating, with RMS Converter (see Table 1 for bandwidths)
Common-Mode Range: ±1500 V operating and without instrument damage
Common-Mode Rejection Ratio: DC and at 60 Hz: infinite; at 1 kHz: -60 dB
Input Impedance (Common-Mode): Essentially infinite for all input ranges
Offset: Initial: ±0.04% of full scale; vs. Temperature: ±20 ppm/°C; vs. Time: ±20
ppm/month
Gain Accuracy: ±0.05% of full scale typical, following calibration*
Gain Stability (all ranges except 5 A): vs. Temperature: ±50 ppm/°C; vs. Time: ±20
ppm/month
Gain Stability (5-A range only): vs. Temperature: ±500 ppm/°C; vs. Time: ±20
ppm/month
(cont'd)
*Initial (uncalibrated) inaccuracy may be as great as ±0.1% of full scale for all input ranges
except 5 A, for which it may be as great as ±0.5%. Maximum error that could occur upon
replacement of a Model 10A68-2 not followed by calibrationis ±0.2% of full scale (±1% for 5-A
range).
MODEL10A68-2
DUALACRMS
C
ONDITIONERCARD
GENERALDESCRIPTIONANDSPECIFICATIONS10A68-2.1
SYSTEM10
ANALOGINPUTCARDS
1.E.2.10A68-2
Product specificaties
| Merk: | Daytronic |
| Categorie: | Niet gecategoriseerd |
| Model: | 10A68-2 |
Heb je hulp nodig?
Als je hulp nodig hebt met Daytronic 10A68-2 stel dan hieronder een vraag en andere gebruikers zullen je antwoorden
Handleiding Niet gecategoriseerd Daytronic
25 Januari 2024
25 Januari 2024
25 Januari 2024
25 Januari 2024
25 Januari 2024
25 Januari 2024
25 Januari 2024
25 Januari 2024
25 Januari 2024
25 Januari 2024
Handleiding Niet gecategoriseerd
Nieuwste handleidingen voor Niet gecategoriseerd
26 Augustus 2026
26 Augustus 2026
26 Augustus 2026
26 Augustus 2026
26 Augustus 2026
26 Augustus 2026
26 Augustus 2026
26 Augustus 2026
26 Augustus 2026
26 Augustus 2026