Bekijk gratis de handleiding van Daytronic 10A73-4 (14 pagina’s), behorend tot de categorie Niet gecategoriseerd. Deze gids werd als nuttig beoordeeld door 33 mensen en kreeg gemiddeld 4.4 sterren uit 5 reviews. Heb je een vraag over Daytronic 10A73-4 of wil je andere gebruikers van dit product iets vragen? Stel een vraag
Pagina 1/14

1GENERALDESCRIPTIONANDSPECIFICATIONS
Intended primarily for stress analysis applications, the Model 10A73-4 receives and
conditions up to
four independent inputs from one-arm (1/4-bridge), two-arm (1/2-
bridge), and four-arm (full-bridge)
strain gage configurations attached directly to the
stressed material.
If you are measuring strain by means of a one-arm (1/4-bridge) strain gage configura-
tion, BE SURE TO STUDY SECTION 4 FOR A DISCUSSION OF “LEAD WIRE AND NON-
LINEARITY EFFECTS WITH QUARTER-BRIDGE STRAIN GAGE CONFIGURATION.”
Unless all four inputs originate from full-bridge transducers, the 10A73-4 requires
external completion of each strain gage input to
full bridge, by means of one of the fol-
lowing:
•aSystem 10 Bridge Completion Connector—one of the following models:
—
10QBC-4-120 for 2- or 3-wire 120-Ωquarter-bridgeconnections
—
10QBC-4-350 for 2- or 3-wire 350-Ωquarter-bridgeconnections
—
10QBC-4-1K for 2- or 3-wire1000-Ωquarter-bridge connections
—
10HBC-4 for 4- or 6-wire half-bridge connections
—
10FBC-4 for 5-wire full-bridge connections
•a Daytronic Model 10CJB-4 Quad Bridge Completion Card(described in Sec-
tion 2.b)
1
;
•equivalent circuitry supplied by the user
2
When the 10A73-4 is used with one of the System 10 Bridge Completion Connectors
listed above, the excitation level is
fixed at 5 V-DC. In the absence of a Bridge Com-
pletion Connector—whether or not the
Model 10CJB-4is used—the 10A73-4 provides
selectable low-level excitation (1, 2, or 5 V-DC) to help reduce gage heating effects in
materials with low thermal conductivity.
3
However, the selection of excitation level for
the 10A73-4 is not “per channel”;
the same level applies to all four 10A73-4 channels.
See Section 3.a for details.
As explained in Section 3.b, the 10A73-4 can be quickly calibrated either through
“calculated” calibration—involving application of the
MVVcommand—or through a
convenient shunt technique. Unless a System 10 Bridge Completion Connector is
being used with the 10A73-4, all four shunt resistors may be switched in and out
simultaneously by means of logic-level inputs through the rear I/O CONNECTOR.
1
The 10CJB-4 contains a switch for easy selection of the common excitation level.
2
If you wish to provide your own bridge-completion circuitry, you should contact the factory for
instructions on connections and calibration.
3
A special modification must be made to the 10A73-4 card to allow selectable excitation when a
System 10 Bridge Completion Connector is attached.
MODEL10A73-4
1/2 & 1/4 BRIDGESTRAINGAGE
CONDITIONERCARD
GENERALDESCRIPTIONANDSPECIFICATIONS10A73-4.1
1.E.2.10A73-4
SYSTEM10
ANALOGINPUTCARDS
Product specificaties
| Merk: | Daytronic |
| Categorie: | Niet gecategoriseerd |
| Model: | 10A73-4 |
Heb je hulp nodig?
Als je hulp nodig hebt met Daytronic 10A73-4 stel dan hieronder een vraag en andere gebruikers zullen je antwoorden
Handleiding Niet gecategoriseerd Daytronic
25 Januari 2024
25 Januari 2024
25 Januari 2024
25 Januari 2024
25 Januari 2024
25 Januari 2024
25 Januari 2024
25 Januari 2024
25 Januari 2024
25 Januari 2024
Handleiding Niet gecategoriseerd
Nieuwste handleidingen voor Niet gecategoriseerd
26 Augustus 2026
26 Augustus 2026
26 Augustus 2026
26 Augustus 2026
26 Augustus 2026
26 Augustus 2026
26 Augustus 2026
26 Augustus 2026
26 Augustus 2026
26 Augustus 2026