Bekijk gratis de handleiding van Daytronic 10A69-4 (6 pagina’s), behorend tot de categorie Niet gecategoriseerd. Deze gids werd als nuttig beoordeeld door 19 mensen en kreeg gemiddeld 4.7 sterren uit 2 reviews. Heb je een vraag over Daytronic 10A69-4 of wil je andere gebruikers van dit product iets vragen? Stel een vraag

Pagina 1/6
1GENERALDESCRIPTIONANDSPECIFICATIONS
The Model 10A69-4 accurately measures the true RMS amplitude up to four indepen-
dent
external analog AC signals, mixed as desired, from a wide range of voltage
sources.
Since the 10A69-4 does not provide excitation, the voltage source must supply its own
power supply
, if required (as is the case with many accelerometers, when used to
measure the noise level or vibration factor of a dynamic process). The three
millivolt
ranges (50, 100, and 200 mV-AC) are intended primarily for current measurement by
means of an external shunt.
ADDITIONAL10A69-4 SPECIFICATIONS
Transducer Types
: AC signal sources (voltage only)
Input Voltage and Current Ranges: See Table 1, below, for allowable full-scale input
ranges (given for each range are the corresponding
normal-mode input impedance
(z
i
), maximum normal-mode input voltage with no instrument damage (V
max
), and
bandwidth flat to 0.2%, 1%, and 3 dB, respectively); automatically selected—on an
individual channel basis—when the channel is configured; for the System 10 channel
“type” codes assigned to 10A69-4 data channels, see Table 2, below
Amplifier (per channel): Selectable gain, followed by high-pass filter with RMSCon-
verter (see Table 1 for bandwidths)
Common-Mode Range: ±1000 V operating and without instrument damage
Common-Mode Rejection Ratio: DC: infinite; at 60 Hz: -60 dB
Input Impedance (Common-Mode): Infinite at DC shunted by 100 pF
Offset: Initial: ±0.02% of full scale; vs. Temperature: ±20 ppm/°C; vs. Time: ±50
ppm/month
Gain Accuracy: ±0.02% of full scale typical, following calibration*
Gain Stability: vs. Temperature: ±50 ppm/°C; vs. Time: ±50 ppm/month
Filter (per channel):3-pole modified Butterworth; 3 dB down at 10 Hz; 60 dB down at
50 Hz
Step-Response Settling Time (Full-Scale Output):
To 1% of final value: 100 msec
To 0.1% of final value: 150 msec
To 0.02% of final value: 600 msec
Auxiliary Outputs to Mainframe Wire-Wrap Pins: None
*Initial (uncalibrated) inaccuracy may be as great as ±0.2% of full scale. Maximum error that
could occur upon replacement of a Model 10A69-4 not followed by calibrationis ±0.4% of full
scale.
MODEL10A69-4
QUADACRMS
C
ONDITIONERCARD
GENERALDESCRIPTIONANDSPECIFICATIONS10A69-4.1
SYSTEM10
ANALOGINPUTCARDS
1.E.2.10A69-4

Beoordeel deze handleiding

4.7/5 (2 Beoordelingen)

Product specificaties

Merk: Daytronic
Categorie: Niet gecategoriseerd
Model: 10A69-4

Heb je hulp nodig?

Als je hulp nodig hebt met Daytronic 10A69-4 stel dan hieronder een vraag en andere gebruikers zullen je antwoorden




Handleiding Niet gecategoriseerd Daytronic

Handleiding Niet gecategoriseerd

Nieuwste handleidingen voor Niet gecategoriseerd