Bekijk gratis de handleiding van Daytronic 10A69-4 (6 pagina’s), behorend tot de categorie Niet gecategoriseerd. Deze gids werd als nuttig beoordeeld door 19 mensen en kreeg gemiddeld 4.7 sterren uit 2 reviews. Heb je een vraag over Daytronic 10A69-4 of wil je andere gebruikers van dit product iets vragen? Stel een vraag
Pagina 1/6

1GENERALDESCRIPTIONANDSPECIFICATIONS
The Model 10A69-4 accurately measures the true RMS amplitude up to four indepen-
dent
external analog AC signals, mixed as desired, from a wide range of voltage
sources.
Since the 10A69-4 does not provide excitation, the voltage source must supply its own
power supply
, if required (as is the case with many accelerometers, when used to
measure the noise level or vibration factor of a dynamic process). The three
millivolt
ranges (50, 100, and 200 mV-AC) are intended primarily for current measurement by
means of an external shunt.
ADDITIONAL10A69-4 SPECIFICATIONS
Transducer Types
: AC signal sources (voltage only)
Input Voltage and Current Ranges: See Table 1, below, for allowable full-scale input
ranges (given for each range are the corresponding
normal-mode input impedance
(z
i
), maximum normal-mode input voltage with no instrument damage (V
max
), and
bandwidth flat to 0.2%, 1%, and 3 dB, respectively); automatically selected—on an
individual channel basis—when the channel is configured; for the System 10 channel
“type” codes assigned to 10A69-4 data channels, see Table 2, below
Amplifier (per channel): Selectable gain, followed by high-pass filter with RMSCon-
verter (see Table 1 for bandwidths)
Common-Mode Range: ±1000 V operating and without instrument damage
Common-Mode Rejection Ratio: DC: infinite; at 60 Hz: -60 dB
Input Impedance (Common-Mode): Infinite at DC shunted by 100 pF
Offset: Initial: ±0.02% of full scale; vs. Temperature: ±20 ppm/°C; vs. Time: ±50
ppm/month
Gain Accuracy: ±0.02% of full scale typical, following calibration*
Gain Stability: vs. Temperature: ±50 ppm/°C; vs. Time: ±50 ppm/month
Filter (per channel):3-pole modified Butterworth; 3 dB down at 10 Hz; 60 dB down at
50 Hz
Step-Response Settling Time (Full-Scale Output):
To 1% of final value: 100 msec
To 0.1% of final value: 150 msec
To 0.02% of final value: 600 msec
Auxiliary Outputs to Mainframe Wire-Wrap Pins: None
*Initial (uncalibrated) inaccuracy may be as great as ±0.2% of full scale. Maximum error that
could occur upon replacement of a Model 10A69-4 not followed by calibrationis ±0.4% of full
scale.
MODEL10A69-4
QUADACRMS
C
ONDITIONERCARD
GENERALDESCRIPTIONANDSPECIFICATIONS10A69-4.1
SYSTEM10
ANALOGINPUTCARDS
1.E.2.10A69-4
Product specificaties
| Merk: | Daytronic |
| Categorie: | Niet gecategoriseerd |
| Model: | 10A69-4 |
Heb je hulp nodig?
Als je hulp nodig hebt met Daytronic 10A69-4 stel dan hieronder een vraag en andere gebruikers zullen je antwoorden
Handleiding Niet gecategoriseerd Daytronic
25 Januari 2024
25 Januari 2024
25 Januari 2024
25 Januari 2024
25 Januari 2024
25 Januari 2024
25 Januari 2024
25 Januari 2024
25 Januari 2024
25 Januari 2024
Handleiding Niet gecategoriseerd
Nieuwste handleidingen voor Niet gecategoriseerd
26 Augustus 2026
26 Augustus 2026
26 Augustus 2026
26 Augustus 2026
26 Augustus 2026
26 Augustus 2026
26 Augustus 2026
26 Augustus 2026
26 Augustus 2026
26 Augustus 2026