TDE Instruments DPM36-P Handleiding


Pagina's 8 Meetapparatuur

Bekijk de handleiding van TDE Instruments DPM36-P gratis online of download de PDF. 11 mensen vonden deze handleiding nuttig. Vraag over dit model? Stel een vraag

Digalox
®
DPM36-P
Instruction manual (Rev-2025-03)
Panel meter for 30V/100mA AC/DC
Package contents:Panel meter Digalox
®
DPM36, 5 jumpers,
2 instruction manuals (EN+DE)
1.Safety instructions
•Read instruction manual carefully before operating the device! Keep for later
reference.
•Mounting and installation must be carried out by suitably qualified and competent
persons only.
•WARNING: When working on the device hazardous voltages must not be
connected to the device! The jumper terminals are not isolated from the
measuring circuit.
•DANGER: When the USB port is connected, only voltages less than 50 V
may be present at the measurement inputs. The USB port is not isolated
from the measuring circuit.
•The device is not intended to protect persons or facilities against harm. Specific
devices must be used to guarantee safety (protection relays, cut-off switches,
etc.).
•When connecting switches to the jumper terminals, only switches must be used
whose isolation voltage is at least twice the maximum occurring measurement
voltage. For example, when measuring 307V DC switches must be isolated for at
least 607V.
•Do not use the device in the presence of explosive or flammable substances!
•All cables carrying hazardous voltages must be secured with external separators.
2.Meaning of symbols
General warning sign
(Attention, observe the documentation!)
----- 1 / 8 -----

Beoordeel deze handleiding


Product specificaties

Merk: TDE Instruments
Categorie: Meetapparatuur
Model: DPM36-P

Heb je hulp nodig?

Als je hulp nodig hebt met TDE Instruments DPM36-P stel dan hieronder een vraag en andere gebruikers zullen je antwoorden