Sonel MRU-120 Handleiding
Sonel Meetapparatuur MRU-120
Bekijk gratis de handleiding van Sonel MRU-120 (2 pagina’s), behorend tot de categorie Meetapparatuur. Deze gids werd als nuttig beoordeeld door 56 mensen en kreeg gemiddeld 4.7 sterren uit 9 reviews. Heb je een vraag over Sonel MRU-120 of wil je andere gebruikers van dit product iets vragen? Stel een vraag
Pagina 1/2

Enabling auto-zeroing
Disabling auto-zeroing
Sonel MRU-120
Earth Resistance Meter QuickStart●
v1.00 | 26.05.2020
Connect the meter to the measured wire.
Run the measurement using START button.
2P
In order to eliminate the inuence of the resistance of the test leads
over the result of the measurement, its compensation (auto-zeroing)
has to be done.
Press START.
Using button enable mode.F1 AUTOZERO
Short-circuit the test leads.
Press START.
Using button enable mode.F1 AUTOZERO
Separate the test leads.
It is sufcient to realize compensation once for the given test leads.
It is also remembered once the meter has been turned off, until the
next successful auto-reset procedure.
First steps
Turn on the meter
Select the method and connect
Congure
Obtain the result
Measurements
The meter is designed for measurements at interference voltages which do not exceed 24 V
for RE measurements and 3 V for RCONTmeasurements. The voltage is measured up to 100 V,
but above 40 V is indicated as dangerous. The meter must not be connected to voltages ex-
ceeding 100 V.
UN>24V!
The voltage on the mea-
surement points exceeds
24 V but is lower than
40 V. The measurement
is blocked.
NOISE!
The value of the interfering
signal is too high. The result
may be distorted by additional
uncertainty.
LIMIT!
The uncertainty of the
electrode resistance
>30%. Uncertainties
calculated on the basis of
the measured values.
UN>40V!
and a contin-
uous
sonic signal
The voltage on the
measurement points
exceeds 40 V. The mea-
surement is blocked.
R>19,99kΩ
RE>19,99kΩ
RS>19,9kΩ
RH>19,9kΩ
ρ>999kΩm
Measurement range
exceeded. I
L>max
Excessive interfering
current, the measurement
error may exceed the
basic error.
Wire continuity measurement
Product specificaties
| Merk: | Sonel |
| Categorie: | Meetapparatuur |
| Model: | MRU-120 |
Heb je hulp nodig?
Als je hulp nodig hebt met Sonel MRU-120 stel dan hieronder een vraag en andere gebruikers zullen je antwoorden
Handleiding Meetapparatuur Sonel
24 December 2025
25 Augustus 2025
19 April 2025
9 December 2024
9 December 2024
9 December 2024
18 November 2024
19 Juli 2024
20 Maart 2024
12 Februari 2024
Handleiding Meetapparatuur
Nieuwste handleidingen voor Meetapparatuur
10 Maart 2026
10 Maart 2026
10 Maart 2026
9 Maart 2026
9 Maart 2026
9 Maart 2026
9 Maart 2026
9 Maart 2026
9 Maart 2026
6 Maart 2026