Bekijk gratis de handleiding van PCE Instruments PCE-PME 250 (8 pagina’s), behorend tot de categorie Meetapparatuur. Deze gids werd als nuttig beoordeeld door 2 mensen en kreeg gemiddeld 4.6 sterren uit 1.5 reviews. Heb je een vraag over PCE Instruments PCE-PME 250 of wil je andere gebruikers van dit product iets vragen? Stel een vraag
Pagina 1/8

User manuals in various languages (français, italiano, español, português, nederlands,
türk, polski) can be found via our product search on:
www.pce-instruments.com
PCE-PME Series
USER MANUAL
ENGLISH
DIGITAL MANOMETER

2
Widely used to measure fan and blower pressure, lter resistance, air speed, furnace pressure, orice
plate dierential pressure, etc. lt is also used for controlling the air to gas ratio and the automatic valve in
combustion process, and monitoring blood pressure and respiratory pressure in medical care equipment.
»Measurement mode switching: Short press button 1 to change between differential pressure /
Maximum (MAX) / Minimum (MIN) / Average (AVG).
Long press button 1 to exit.
»Air pressure unit switching: Short press button 2 to change between barmbar / Pa / Kgf/cm2 / mmHg /
cmH2O / Ozf/in2 / Psi / inHg / inH2O / ftH2O / HPa / Pa.
Long press button 2 to change directly to Pa.
»Air pressure data hold/not hold: short press button 3.
»Buzzer alarm on/off: long press button 3.
Low pressure chamber High pressure chamber
A. SCOPE OF APPLICATION
B. KEY DESCRIPTION AND FUNCTION OPERATION
DEVICE DESCRIPTION

3
»Time-date switch: short press button 4.
»Barometric pressure zero calibration: long press button 4.
»Power on: short press button 5.
»Power off: long press button 5.
»Automatic shutdown function;
On: Automatie shutdown time setting: Computer application > settings > fill in the automatic shutdown
time > check the box behind the automatic shutdown > click an the upload button in this area.
O (this time): lf the automatic shutdown lunction has been opened through the computer application,
short press button 5 to display „uoff“, indicating that the automatic shutdown function is only temporarily
canceled, and the automatic shutdown function is still an alter next start-up.
O (permanently): Computer application > settings > da not check the box behind automatic shutdown >
click an this area to upload button icon.
»Record data (air pressure value, air pressure unit, time and date) function:
On: short press button 6 or computer application > settings > fill in the storage interval (1s ... 9999s) > turn
on the storage switch > Click on this area to upload the button icon;
O:
If the barometric data recording lunction has been turned on, short press button 6, or go to computer
application > settings > close the storage switch > click on the upload button icon in this area:
Clear: Press and hold button 6 or go to computer application > click on the delete storage icon;
View storage space: Computer application > settings > saved notes, the maximum number of notes that
can be stored is 15996;
View recorded data: Computer application > click on the download button icon;
Note: The more recorded data, the slower the data export time, please check the progress bar at the bot-
tom right of the computer application when exporting.
»Real-time monitoring function of pressure dala graph:
lnterval setting: Computer application > settings > lill in the measurement interval (unit: seconds);
»Operating instructions lor computer application:
Computer application > help (H) > help (H) > click „yes“ to open the help document.
»Buzzer alarm function:
On: Press and hold button 3 or go to computer application > settings > set the high alarm value > select all
the numbers including the plus and minus signs behind > press DELETE on the computer keyboard > re-
enter the alarm value, such as +2 (unit: Pa) > set the low alarm value > select all the numbers including the
plus and minus signs behind -> press DELETE on the computer keyboard -> re-enter the alarm value, such
as -3 (unit: Pa) > turn on the alarm switch > click on the upload button icon in this area;
O: Computer application > settings > turn off the alarm switch > click on the upload button icon in this
area or press and hold button 3 to turn off.
Note: The range of high and low alarm values that can be set varies depending on the pressure gauge‘s
range.
Product specificaties
Merk: | PCE Instruments |
Categorie: | Meetapparatuur |
Model: | PCE-PME 250 |
Heb je hulp nodig?
Als je hulp nodig hebt met PCE Instruments PCE-PME 250 stel dan hieronder een vraag en andere gebruikers zullen je antwoorden
Handleiding Meetapparatuur PCE Instruments
16 Juni 2025
15 Juni 2025
15 Juni 2025
25 Mei 2025
22 Mei 2025
22 Mei 2025
22 Mei 2025
14 Mei 2025
14 Mei 2025
14 Mei 2025
Handleiding Meetapparatuur
- Testboy
- Motrona
- Aktobis
- Bresser
- Lanaform
- Shimano
- Kyoritsu
- Techno Line
- Hazet
- Schneider
- Soler And Palau
- Hama
- SecuTech
- Eizo
- Ermenrich
Nieuwste handleidingen voor Meetapparatuur
16 Juni 2025
15 Juni 2025
15 Juni 2025
15 Juni 2025
15 Juni 2025
15 Juni 2025
15 Juni 2025
15 Juni 2025
15 Juni 2025
12 Juni 2025