Mean Well GSM60B05-P1J Handleiding


Bekijk gratis de handleiding van Mean Well GSM60B05-P1J (7 pagina’s), behorend tot de categorie Niet gecategoriseerd. Deze gids werd als nuttig beoordeeld door 25 mensen en kreeg gemiddeld 4.3 sterren uit 13 reviews. Heb je een vraag over Mean Well GSM60B05-P1J of wil je andere gebruikers van dit product iets vragen? Stel een vraag

Pagina 1/7
Model::GSM60B5-P1J Test Report 1 / 7
Test Report: GSM60B5-P1J
60W AC- DC Reliable Green Medical Adaptor
DESIGN VERIFY TEST
O F Tutput unction est
Input Function Test
Protection Function Test
C F Tontrol unction est
Component Stress Test
SAFETY & . E.M.C TEST
Safety Test
E.M.C. Test
RELIABILITY TEST
ENVIRONMENT TEST
60W AC-DC Rel iable Green Medical Adaptor
GSM60B series
Model::GSM60B5-P1J Test Report 2 7 /
DESIGN VERIFY TEST
OUTPUT FUNCTION TEST
NO
SPECIFICATION
TEST CONDITION
RESULT
VERDICT
1 RIPPLE & NOISE V180
mVp-p (Max )
I/P 230VAC
O/PFULL LOAD
Ta25
V1
41.0
mVp-p (Max )
P
2 OUTPUT VOLTAGE
TOLERANCE
V1 -5 +5 %~ % (Max) I/P80 VAC / 264 VAC
O/PFULL/ MIN LOAD
Ta25
V1
- 2.00
%~
2.24
%
P
3 LINE REGULATION V1 -1 +1 %~ % (Max)
I/P100 VAC ~ 264 VAC
O/PFULL LOAD
Ta25
V1
0
%~
0 .12
%
P
4 LOAD REGULATION
V1
-5 +5 %~ % (Max)
I/P
230 VAC
O/PFULL ~MIN LOAD
Ta25
V1
- 2.00
%~
2.24
%
P
5 SET UP TIME
230VAC
1000
ms (Max)
I/P 230 VAC
I/P 115 VAC
O/PFULL LOAD
Ta
25
230VAC/
422
ms
P
115VAC
1500 ms(Max) 115VAC/ 678 ms
6 RISE TIME
230VAC
30
ms (Max)
I/P
230 VAC
I/P 115 VAC
O/PFULL LOAD
Ta25
230VAC/
5.2
ms
P
115VAC 30 ms (Max) 115VAC/ 11.9 ms
7 HOLD UP TIME
230VAC
50
ms (TYP)
I/P 230 VAC
I/P 115 VAC
O/PFULL LOAD
Ta
25
230VAC/
144
ms
P
115VAC
16 ms (TYP) 115VAC/ 25.6 ms
8 OVER/UNDERSHOOT
TEST
< + 5%
I/P
230 VAC
O/PFULL LOAD
Ta25
TEST
0 %
P
9 DYNAMIC LOAD V1 1 -p 000 mVp I/P 230 VAC
(1).O/PFULL /Min LOAD 90%DUTY/
1KHZ
(2).O/PFULL /Min LOAD 90%DUTY/
3 KHZ
(3).O/PFULL /Min LOAD 90%DUTY/
5 KHZ
(4).O/PFULL /Min LOAD 50%DUTY/
120HZ
Ta
25
(1) 680
(2) 652
(3) 652
(4) 674
mVp-p
mVp-p
mVp-p
mVp-p
P
INPUT FUNCTION TEST
NO
TEST ITEM
SPECIFICATION
TEST CONDITION
RESULT
VERDICT
60W AC-DC Rel iable Green Medical Adaptor
GSM60B series
Model::GSM60B5-P1J Test Report 3 7 /
1 INPUT VOLTAGE RANGE 80VAC~264 VAC I/PTESTING
O/PFULL LOAD
Ta25
71.2 V~264V
P
I/P
- - V LOW LINE 3V= 77
- HIGH LINE+15%=300 V
O/PFULL/MIN LOAD
ON30 Sec . OFF30 Sec 10MIN
( AC POWER ON/OFF NO DAMAGE )
TEST OK
2 INPUT FREQUENCY
RANGE
47HZ ~63 HZ
NO DAMAGE OSC
I/P
80 VAC ~ 264 VAC
O/PFULL~MIN LOAD
Ta25
TEST
OK
P
3 EFFICIENCY 81.5% (TYP) I/P 230 VAC
O/PFULL LOAD
Ta
25
85.87 %
P
4 INPUT CURRENT
230V/
1
A (TYP)
I/P
230 VAC
I/P 115 VAC
O/PFULL LOAD
Ta25
I =
0 .29
A/ 230 VAC
P
115V/ 1.4 A (TYP) I = 0.52 A/ 115 VAC
5 INRUSH CURRENT 230V/
115V/
60
30
A (TYP)
A (TYP)
I/P 230 VAC
I/P 115 VAC
O/PFULL LOAD
Ta25
I =
I=
41.523
20.875
A/ 230 VAC
A/ 230 VAC
P
COLD START
6 LEAKAGE CURRENT 50 u A / 264VAC I/P 264 VAC
O/PMin LOAD
Ta25
FOR PATIENT
34.4μA
P
7 NO LOAD CONSUMPTION
PS-ON SHORT < 1 W 0.
I/P
24 0VAC
O/PNO LOAD
Ta
25
<
0.0499 W
P
PROTECTION FUNCTION TEST
NO
TEST ITEM
SPECIFICATION
TEST CONDITION
RESULT
VERDICT
1 OVER LOAD PROTECTION % ~ % 105 160 I/P 230 VAC
I/P 115 VAC
O/PTESTING
Ta25
128.8 %/ 230 VAC
125.9 %/ 115 VAC
Protection type : Hiccup mode,
recovers automatically after fault
condition is removed
P
2 OVER VOLTAGE PROTECTION
CH1
5.2 7.0V ~ V
I/P
230 VAC
I/P 115 VAC
O/PMIN LOAD
Ta25
6.12 V/ 230 VAC
6.15 V/ 115 VAC
Protection type : Shut down o/p
voltage, re-power on to recover
P
3 OVER TEMPERATURE PROTECTION SPEC
RTH2>70
NO DAMAGE
I/P 230 VAC
O/PFULL LOAD
O.T.P. Active
Shut down o/p voltage, re power on -
to recover
P


Product specificaties

Merk: Mean Well
Categorie: Niet gecategoriseerd
Model: GSM60B05-P1J
Gewicht: 320 g
Breedte: 50 mm
Diepte: 125 mm
Duurzaamheidscertificaten: RoHS
Aantal per verpakking: 40 stuk(s)
Ingangsspanning: 80 - 264 V
Uitgangsvermogen: 60 W
Output current: 6 A
Spanning: 5 V