Bekijk gratis de handleiding van Labec AA13 Autoclaaf (7 pagina’s), behorend tot de categorie Autoclaven. Deze gids werd als nuttig beoordeeld door 13 mensen en kreeg gemiddeld 4.4 sterren uit 7 reviews. Heb je een vraag over Labec AA13 Autoclaaf of wil je andere gebruikers van dit product iets vragen? Stel een vraag
Pagina 1/7

Laboratory Equipment Pty Ltd Instruction Manual - Autoclave
2
INSTRUCTION MANUAL
Autoclave Floor Model +135ºC
MODELS: AA13, AA16, AA18, AA20

Introduction
Congratulations on the choice of an Australian made quality product. Labec products are manufactured,
tested and calibrated to meet published standard specifications under our strict quality assurance
guidelines.
This Instruction Manual is for the guidance of operators of Labec Autoclaves and should be read before the
autoclaveisconnectedtotheelectricitysupply.
It is hoped that this manual will supply all the information that the customer should require for satisfactory
operation of this product. If, however, there are any questions that remain unanswered please contact the
ServiceDepartmentviathewebsiteenquirypageandincludetheserialnumberoftheunit.
Unpacking
Remove all packing and protective wrapping from both interior and exterior of the autoclave. Check the
autoclave for any possible transit damage. Ensure all ordered accessories are present. If any physical
damage or shortages are evident, do not discard the packaging material until the autoclave is inspected by
thedistributor,agentormanufacturer.
NOTE:Allclaimsforshortageordamagemustbemadewithinfourteendays(14)fromdelivery.
Subject to our standard published conditions of sale, we have reasonable grounds to believe that we have
ensured, so far as is reasonably practical, that the products listed in our catalogue and brochures have been
designed and constructed so as to be safe and without risk to health when properly installed and used in
their environment by appropriate and trained personnel, and where applicable, in accordance with our
publishedinstructions.
Installation
Electrical
This equipment must be tagged and tested according to AS/NZS3760:2010 prior to use and
thereafterevery6monthsdependentupontheenvironment.
A qualified electrician should connect the power supply through the conduit at the back of the autoclave.
Access to the terminal block is obtained by removing the right hand side panel of the autoclave. The frame
of the autoclave must be sufficiently earthed using the earth terminal located adjacent to the
terminalblock.
Ensure Autoclave is screwed to bench at rear using attaching lugs to prevent pulling or tipping Autoclave off
benchwheninuse.
It is preferable to locate the autoclave close to a power point and recommended that double adaptors are not
used. Check the total wattage if connecting to multipoint outlets. Check the rating plate for power
requirements. Installation is to be carried out by a qualified electrician in accordance with the power
requirementsoftheproductspecifications.
SeeAppendixBattached.
Plumbing
Connectallplumbinginaccordancewithlocalauthority’sregulations.
SeeAppendixCattached
1LaboratoryEquipmentPtyLtdInstructionManualAutoclaveAA13,16,18,20Ver2.512013

Location
Select a location free from draught and away from direct sunlight or other heat source. Proximity to a drain
for:
1. Wastefromthemanuallyoperateddrain.
2. Theoutletsservicingthepressurereleasevalveandairvent
3. Adjustablesteamreleasevalveatthebottomrighthandrearofcabinet.
General
The LABEC Automatic Autoclave is designed and constructed to meet the requirements of AS1210 of the
Pressure Vessel Code and has been approved by N.S.W. Workcover Authority, to operate at pressures up
to 215 K.P.A. Regulations require that on installation the autoclave be examined by a “Competent Person” or
licensed boiler inspector and this certificate of inspection be displayed near the autoclave and renewed and
inspected annually. For more details contact your local equivalent of Workcover NSW or contact Labec for
moreinformation.
ApprovalNumbers
NSW
QLD
SA
WA
TAS
AA9
6766P87
WA94
AA13/AA16
674P90
BV18716
PV86636/7
WA9489
TA4946
AA18/AA20
673P90
BV18717
PV86636/7
WA9490
TA4947
Heating
Heating of the autoclave is by means of stainless tubular element located in the base of the autoclave
vessel.
DescriptionofControls
SeeAppendixAAttached
TemperatureControl
Labec autoclaves are fitted with solid state proportional action temperature controllers which operate from a
sensitive thermocouple or Rtd inserted in the working space of the chamber. The controller has been
calibrated at 121°C and before any adjustments to temperature settings are made allow the autoclave
temperature to stabilise for at least one half hour. For the digital controller fitted please read the enclosed
operating instructions when setting the controller temperature. The timer will automatically start the
countdown process once the temperature reaches the set value. (OP4 indicates the timer has
initiated).
ImportantSafetyInstructions
Caution
PleaseobservethefollowingsafetymeasuresbeforeusingyourLABECequipment.
1. Ensure pressure inside the autoclave is reading zero KPA on the pressure gauge before opening the
lid.
2. These autoclaves are and under no circumstances should inflammable, NOT FLAME PROOF
combustibleorexplosivematerialbeplacedintheautoclave.
3. Low ignition temperature materials and those materials which give off inflammable or explosive
vaporsshouldnotbeplacedintheautoclave.
4. Avoidheatingsubstanceswhichgiveoffcorrosivevapor.
5. Users are advised of the dangers of heating combustible materials. The manufacturer can
recommendspecialtypesofelementswhichwillprohibittheautoclavestemperature
6. Observe those rules pertaining to wiring and installation of electrical appliances as recommended by
2LaboratoryEquipmentPtyLtdInstructionManualAutoclaveAA13,16,18,20Ver2.512013