Bekijk gratis de handleiding van Gossen Metrawatt PROFITEST PRCD M512R (16 pagina’s), behorend tot de categorie Niet gecategoriseerd. Deze gids werd als nuttig beoordeeld door 28 mensen en kreeg gemiddeld 4.5 sterren uit 14.5 reviews. Heb je een vraag over Gossen Metrawatt PROFITEST PRCD M512R of wil je andere gebruikers van dit product iets vragen? Stel een vraag
Pagina 1/16

Operating Instructions
PROFITEST⏐PRCD
Adapter for Standards Compliant Testing of PRCDs
by Simulating Faults 3-349-797-15
7/4.19
Important
Read carefully before use!
Keep on file for future reference!
Please observe the manufacturer‘s details
on the devices under test!

2 GMC-I Messtechnik GmbH
Connections Overview
1 Coupling plug to the outlet of a
3-phase PRCD
2 Earthing contact plug to the outlet of a
single-phase PRCD
3 Probe socket for contacting the on/off
key on the PRCD
4 Earthing contact socket to the inlet of a
single-phase PRCD
5 Loop for measuring protective conductor
current with a current clamp transformer
6 CEE socket to the inlet of a 3-phase
PRCD
7 Mains power cable with earthing contact plug
8 3-phase mains power cable with
CEE outlet (1P+N+PE 16 A-6h)
9 PRCD inlet sockets connected in parallel
to sockets 4 and 6 – for connecting the
PROFITEST MXTRA during protective con-
ductor and insulation measurements
10 PRCD outlet sockets connected in paral-
lel to plugs 1 and 2 for connecting the
PROFITEST MXTRA during protective con-
ductor and insulation measurements
11 Five replaceable fuses
Instrument Overview
8
7
6
4
1
2
10
3
9
Mains Connection
To PRCD Outlet
To PRCD Inlet ( )PRCD IN
5
11
( )PRCD OUT
Fault Selection Mains disconnect, 1~OFF/3~OFF
– For RISO measurement
Mains power on, 1~ON/3~ON
– For fault simulation
– For RLO measurement
Phase LEDs
Phase Interrupt
Reversed Wires
Orange positions: testing at single-phase PRCDs
Green positions: testing at 3-phase PRCDs
Mains
– For tripping test with I
Nom
and measurement of time to trip
Connection
PROFITEST⏐PRCD Operating and Connections Overview

GMC-I Messtechnik GmbH 3
Meanings of Symbols on the Instrument
300 V CAT II
Maximum permissible voltage and
measuring category between con-
nections and ground
Warning concerning a point of
danger
(attention:
observe documentation!)
Indicates European Conformity
This device may not be disposed
of with the trash.
Further information regarding the
WEEE mark can be accessed on
the Internet at www.gossenme-
trawatt.com by entering the
search term “WEEE”.
Meanings of Symbols in the Operating Instructions
LED L1, L2, L3, N or PE at test adapter
LED on LED off
PRCD LED at the device under test
LED on LED off
Scope of Delivery
1 Test adapter
1 Probe cable with plug-in test probe
1 Set of operating instructions
Table of Contents Page
1 Safety Precautions .................................... 4
2 Applications .............................................. 5
3 Initial Start-Up ........................................... 6
3.1 Testing the LEDs ......................................... 6
3.2 Mains Connection ....................................... 6
3.3 Connecting the PRCD .................................. 7
4 Measurements with the
PROFITEST MXTRA .................................... 7
4.1
Measuring Protective Conductor Resistance (Rlo) ....... 7
4.2
Measuring Insulation Resistance (Riso) ............. 8
4.3
Tripping Test with Nominal Residual Current and
Measurement of Time to Trip ...............................8
4.4 Varistor Test (PRCD-K) ................................. 9
5 Protective Conductor Current Measurement 9
6 Fault Simulation ...................................... 10
6.1 PRCD-S (single-phase) ............................. 10
6.1.1 Simulated Interruption ....................................... 10
6.1.2 Simulated Wire Reversal .................................... 10
6.1.3 Simulation of PE to Phase – PE-UEXT ................. 10
6.1.4 Contacting the “ON” Key at the PRCD with the Probe......11
6.2 PRCD-K (single-phase) ............................. 11
6.2.1 Simulated Interruption ....................................... 11
6.3 PRCD-S (3-phase) .................................... 12
6.3.1 Simulated Interruption ....................................... 12
6.3.2 Simulated Wire Reversal .................................... 13
6.3.3 Simulation of PE to Phase ................................... 13
7 Characteristic Values .............................. 14
8 Maintenance ........................................... 14
8.1 Housing Maintenance ................................ 14
8.2 Safety Checks – Testing in Compliance with
DGUV Provision 3 ...................................... 15
8.2.1
Testing of Protective Conductor Resistance RPE ............15
8.2.2 Testing of Insulation Resistance: ......................... 15
8.2.3 Contact Current Measurement ............................ 16
8.3 Fuse Replacement ..................................... 16
8.4 Return and Environmentally Sound Disposal 16
9
Repair and Replacement Parts Service, Calibration
Center and Rental Instrument Service ................. 16
10 Product Support ...................................... 16
!
PROFITEST⏐PRCD Symbols – Scope of Delivery – Table of Contents
Product specificaties
Merk: | Gossen Metrawatt |
Categorie: | Niet gecategoriseerd |
Model: | PROFITEST PRCD M512R |
Heb je hulp nodig?
Als je hulp nodig hebt met Gossen Metrawatt PROFITEST PRCD M512R stel dan hieronder een vraag en andere gebruikers zullen je antwoorden
Handleiding Niet gecategoriseerd Gossen Metrawatt
16 December 2024
13 November 2024
6 November 2024
5 November 2024
5 November 2024
5 November 2024
5 November 2024
8 April 2024
19 Maart 2024
11 Maart 2024
Handleiding Niet gecategoriseerd
- T.akustik
- Dormakaba
- SureFire
- Purell
- Festool
- Namco Bandai Games
- Polsen
- Meister Craft
- Ventus
- Rooboost
- Pyle Pro
- Dot Line
- 3 Legged Thing
- Miniland
- Buffalo
Nieuwste handleidingen voor Niet gecategoriseerd
15 Juni 2025
15 Juni 2025
15 Juni 2025
15 Juni 2025
15 Juni 2025
15 Juni 2025
15 Juni 2025
15 Juni 2025
15 Juni 2025
15 Juni 2025