Introduction
The Model 80T-150UA Universal Temperature Probe is a self-contained
temperature-to-voltage converter. The probe is designed to provide a
direct temperature reading when it is connected to any high impedance
DMM that is capable of 1 mV resolution and at least a 300-count full
scale readout capability. Output is 1 mV per degree (Celsius or
Fahrenheit). Two switch-selected temperature output scalings are
provided: -50 to +150 °C or -58 to +302 °F. The probe will stand off 60 V
dc or 30 V ac rms (42.4 V peak).
The unit is housed in two separate but attached assemblies: a
temperature probe and a temperature-to-voltage converter. The probe
contains the temperature-sensing element and is electrically connected to
the temperature-to-voltage converter through a 46-inch cable. A three-
position switch on the converter acts as a power switch and is used for
selecting Celsius or Fahrenheit scaling for the output. Two banana plugs
with standard 0.75-inch spacing are provided for connecting the 80T-
150UA to the DMM.
Operating power for the 80T-150UA is derived from a standard 9 V
battery. Typically, an alkaline battery provides more than 600 hours of
continuous operation before replacement is necessary. An OFF switch is
provided on the temperature-to-voltage converter to allow battery
conservation when the unit is not in use. In addition, the OFF position of
the power switch allows the battery condition to be determined via the
external DMM.
Temperature is measured by exposing the probe tip directly to the
material to be measured (non-corrosive liquid, gas, or solid). A direct
temperature reading is displayed on the DMM.
Symbols
XWarning: Risk of electric shock. Refer to the manual for
information.
W Attention: Refer to the manual for information about this
feature.
PN 3208738
February 2008 , Rev.1, 6/09
© 2008-2009 Fluke Corporation. All rights reserved.
All product names are trademarks of their respective companies.
Specifications
The 80T-150UA will achieve rated accuracy when it is used with any
0.25 % DMM that has an input impedance of . ≥1 MΩ
Electrical
Measurement Range: (See Probe Limitations)
-50 to +150 °C
-58 to +302 °F
Accuracy: (Applies for one year after purchase or calibration)
AMBIENT °C ACCURACY
+15 to +35 ° ± °C 1 C from 0 to +100 °C, decreasing
linearly to ± °3 C at -50 and +150 °C
0 to +15 °C and ± °2 C from 0 to +100 °C,
+35 to +50 °C decreasing linearly to ±4 °C at -50
and +150 °C
AMBIENT °F ACCURACY
+59 to +95 ° ± ° °F 1.8 F from +32 to +212 F,
decreasing linearly to ±5.4 °F at -58
and +302 °F
+32 to +59 °
F and ± ° °3.6 F from +32 to +212
F, +95 to +122 °F decreasing linearly to ±7.2 °F at -58
and +302 °F
Sensitivity (80T-150UA output): 1 mV dc / °C or °F
Voltage Standoff: 60 V dc or 30 V ac rms (42.4 V peak)
Settling Time: 9.5 seconds to settle within 2 ° for a 50 ° change
RF Field: The 80T-150UA product is unspecified in environments with
EMC fields greater than 100 mV/meter. ESD performance is specified as
Criteria B per EN61326.
Environmental
Ambient Operating Range for Temperature Module: 0 to +50 °C
(+32 to +122 °F)
Maximum Temperature Probe Body and Cable: +70 °C
(+160 °F) See Probe Limitations
Minimum Temperature Probe Body and Cable:
Minimum Operating Temperature for Cable: -12 °C (+10 °F)
Minimum Operating Temperature for Probe Body: -40 °C (-40 °F)
See Probe Limitations
Storage Temperature for Unit: -40 to +70 °C (-40 to +160 °F)
Humidity: 0 % to 90 % (0 ° °C to +35 C)
0 % to 70 % (+35 °C to +50 °C) (noncondensing)
Altitude: Operating: ≤ ≤3048 m ( 10,000 ft)
Storage: ≤ ≤15240 m ( 50,000 ft)
Application Force: 15 pounds maximum (probe tip to measured surface)
General
Weight: 150 g (5.3 oz)
Overall Length: 53.8 inch (1.36 m)
Battery: Standard 9 V battery (NEDA #1604,6F22,006P)
Battery Life: 600+ hours, typical (Alkaline Battery), 6.5 V minimum
Output Termination: Standard 0.75-inch spaced double banana plug
Probe Material: Glass-filled valox
Probe Size: 0.6 inch maximum diameter
Tip Material: Nickle Plated Copper
Tip Size: 0.31 cm (0.12 inch diameter)
Operating Notes
The following paragraphs are intended to familiarize the operator with
the 80T-150UA. The operator should read these paragraphs before
attempting to operate the probe.
Probe Limitations
The 80T-150UA probe is constructed of a highly durable plastic and is
suitable for measuring the temperature of liquids, gases, and solid
surfaces up to 150 °C. When measuring temperature, observe the
following precautions to prevent damage to the probe:
1. Do not expose the probe end (probe tip plus about 2 inches of the
probe body) to temperatures exceeding +150 °C (+302 °F). The
remainder of the probe body should not be exposed to temperatures
above +70 °C (+160 °F).
2. Do not expose the probe end (probe tip plus about 2 inches of the
probe body) to temperatures below -50 °C (-58 °F). The remainder
of the probe body should not be exposed to temperatures below
-40 °C (-40 °F).
3. For liquid measurements, recommended applications range from
water, lubricants, and fuels to most solvents. Liquids as shallow as
1.27 cm (0.5 inch) can be measured since the temperature sensor is
inch the probe tip. XWWarning
To avoid electrical shock, do not use this instrument when
voltages exceeding 60 V dc or 30 V ac rms (42.4 V peak )
are present. The probe tip is electrically connected to the
output terminals.
WCaution
Long-term exposure of the probe to corrosive
environments will result in pitting and deterioration of the
aluminum probe tip. If equipment is used in a manner not
specified by the manufacturer, the protection provided by
the equipment may be impaired.
Error Sources
When the probe tip is applied to a solid surface, it draws or sinks heat
from the surface. Therefore, if the measured surface has a low mass
(e.g., a transistor case), the indicated temperature may be lower than
the actual temperature.
Similarly, a steady-state error or gradient exists between the measured
surface and the sensing device in the probe tip. This is due to the flow of
heat from the measurement surface to the probe body. The effect of the
steady-state error increases as the differential between ambient and
surface temperature increases.
To determine the actual surface temperature of a device, both the heat-
sinking and steady-state errors must be considered. The correction curve
given in Figure 1 approximates the effect of both error sources on TO-3,
TO-5, and TO-18 transistor cases.


Product specificaties

Merk: Fluke
Categorie: Niet gecategoriseerd
Model: 80T-150UA
Kleur van het product: Black,Red
Gewicht: 150 g
Diepte: 1360 mm
Stroombron: Batterij/Accu
Ondersteuning voor plaatsing: Draagbare multimeter
Accu/Batterij voltage: 6.5 V
Interface: Banaanstekker
Aantal per verpakking: 1 stuk(s)
Lengte: 1360 mm
Temperatuur bij opslag: -40 - 70 °C
Hoogte, in bedrijf: 0 - 3048 m
Bereik temperatuurmeting: -50 - 150 °C
Contact geleider materiaal: Nikkel
Levensduur batterij: 600 uur
Maximale hoogte, in gebruik: 3048 m
Bedrijfstemperatuur (T-T): 0 - 50 °C
Relatieve vochtigheid in bedrijf (V-V): 0 - 90 procent
Type product: Testsonde
Batterijtechnologie: Alkaline
Type batterij: 9V
Temperature measurement range: -58 - 302 °F
Nauwkeurigheid van de temperatuur: 3.6 °C
DC-spanning (max): 60 V
Resolutie (DC voltage): 1 mV
Maximale opslag hoogte: 15240 m
Connectordiameter: 31 mm
Type thermokoppel: Type K

Heb je hulp nodig?

Als je hulp nodig hebt met Fluke 80T-150UA stel dan hieronder een vraag en andere gebruikers zullen je antwoorden




Handleiding Niet gecategoriseerd Fluke

Fluke

Fluke TiX885 Handleiding

7 December 2023
Fluke

Fluke TiX875 Handleiding

7 December 2023
Fluke

Fluke TiX880 Handleiding

7 December 2023
Fluke

Fluke TiX870 Handleiding

7 December 2023
Fluke

Fluke EI-162BN Handleiding

9 Augustus 2023
Fluke

Fluke CNX C3003 Handleiding

8 Augustus 2023
Fluke

Fluke 80i-2010s Handleiding

8 Augustus 2023
Fluke

Fluke TiS20+ Handleiding

8 Augustus 2023
Fluke

Fluke i1010 Handleiding

8 Augustus 2023

Handleiding Niet gecategoriseerd

Nieuwste handleidingen voor Niet gecategoriseerd