Extech 380400 Handleiding


Lees hieronder de đź“– handleiding in het Nederlandse voor Extech 380400 (1 pagina's) in de categorie Niet gecategoriseerd. Deze handleiding was nuttig voor 35 personen en werd door 2 gebruikers gemiddeld met 4.5 sterren beoordeeld

Pagina 1/1
3 - -GB_V1.5 9/19 80400 en
User Manual
Resistance Decade Box
Model 380 0 40
Introduction
Congratulations on your purchase of the Extech 380400 Resistance Decade Box. This device
oers 7 decades of resistance ranges from 1 to over 11MOhms (in 1Ohm steps). Slide switches
allow easy addition and subtraction of resistance values. Binding posts (3) oer simp and le
secure connections. Careful use of this decade box will provide years of reliable service.
Specications
Resistance ranges 1 to 11,111,110 Ohms in 1 Ohm steps
Internal Resistance 0.3 Ohms
Power 0.3W Resistors
Connection Three (3) binding posts
Accuracy 1% Reading) + 0.3 Ohms ±(
Operating conditions Temperature: 32 to 122oF (0 to 50oC) / Humidity: 80%RH <
Dimensions/Weight 5.79 x 4.61 x 1.3" (14.7 x 11.7 x 3.3 Approx. 0.69 lbs. (312 cm) / g)
Operation
Binding Post Connections
The binding posts can be used for connections in several ways:
1. A banana plug can be inserted directly into the posts.
2. Bare wire can be threaded through the post after it has been unscrewed. Once the bare wire is threaded, tighten the posts as
necessary.
3. Alligator clips can be used but use caution not to strip the post threads or plastic post housing.
The resistance output is available on the RED and the BLACK binding posts. The WHITE post is case ground and is typically not
used. Connect the positive lead of the device under test to the RED post. Connect the negative lead to the BLACK post. Use the
WHITE grounding post only if the device under test will be grounded to the 380400 case.
Range Selection
The front panel switches are used to select the resistance that will be available on the RED and BLACK terminals. When a switch 28
is set to the IN position, the value printed above the switch is added to the total resistance available at the posts. When the switch is
set to OUT it is excluded from the total resistance. If all of the switches are set to OUT, the total output resistance zero (+ 0.3Ohms is
internal resistance approx.). –
For example, if the desired output value is 10.5KOhms, set the following switches to the IN position: 10K, 400Ohms and 100Ohms , .
Testing
This device can be used to verify the calibration integrity of MultiMeters, LCR meters, Calibrators, etc. Connect as described above in
the Binding Post Connection section. Set the resistance switches to output the desired resistance. Ensure that the voltage supplied
by the device under test does not overload the 0.3W internal resistor power rating. The device under test should read the value of
resistance selected on the tester. If it does not, the device under test may need calibration, adjustment, or repair.
Copyright - FLIR Systems, Inc. ©2013 2019
All rights reserved including the right of reproduction in whole or in part in any form
www.extech.com


Product specificaties

Merk: Extech
Categorie: Niet gecategoriseerd
Model: 380400

Heb je hulp nodig?

Als je hulp nodig hebt met Extech 380400 stel dan hieronder een vraag en andere gebruikers zullen je antwoorden




Handleiding Niet gecategoriseerd Extech

Handleiding Niet gecategoriseerd

Nieuwste handleidingen voor Niet gecategoriseerd