Bekijk gratis de handleiding van Daytronic 10A35 (4 pagina’s), behorend tot de categorie Niet gecategoriseerd. Deze gids werd als nuttig beoordeeld door 26 mensen en kreeg gemiddeld 4.4 sterren uit 8 reviews. Heb je een vraag over Daytronic 10A35 of wil je andere gebruikers van dit product iets vragen? Stel een vraag
Pagina 1/4

1GENERALDESCRIPTIONANDSPECIFICATIONS
This single-channel, phase-encoded conditioner may be used with incremental
encoder sensors
whose outputs are quadratured to indicate “travel” and “direction of
travel,” for measurement of relative linear or angular displacements over a very wide
range of distances and precisions.
The 10A35 generates a count per 90°phase shift, with a maximum allowable count of
±32700 and a speed limitation of 50 kHz, maximum. To accommodate the largest num-
ber of transducers, this count may be predivided by any integer from 1 to 256, as
explained in Section 3.a. The 10A35 allows run-time “zero indexing” (resetting) and
restarting of a channel’s count reading by the operator or host computer. There are also
input provisions for a user-supplied auxiliary external power supply from 9 to 15 V-DC,
which would allow you to keep the count “alive” during system power shut-down.
NOTE: A System 10 data channel derived from a Model 10A35 always takes a channel
“type” code of
CA, and does not require calibration.
ADDITIONAL10A35 SPECIFICATIONS
Input Signals: ZEROINDEX(RESET), PHASE 1 (SINE), and PHASE 2 (COSINE); each
input is available as a “TTLINPUT” or a “LOW-LEVELINPUT”
Excitation: Nominal 5 V-DC; 50 mA maximum
Accuracy: Based on transducer, ±1 digit
Auxiliary Power Supply: 9 to 15 V-DC
Auxiliary Outputs to Mainframe Wire-Wrap Pins:None
2TRANSDUCERCONNECTIONS
The Model 10A35’s I/O CONNECTOR mates with Daytronic CONDITIONERCONNEC-
TOR No. 60322, shown in Fig. 1.5 (in Manual Section 1.E.1). Standard cable connec-
tions—including connection of an optional “keep-alive” power supply—are given in Fig.
1, below. Table 1 gives standard pin assignments for the I/O Connector.
The 10A35 takes three input signals:
ZEROINDEX(RESET), PHASE 1 (SINE),and
PHASE 2 (COSINE). Each input is available as a “TTLINPUT” or a “LOW-LEVEL
INPUT.” With internal pull-up less than 1 LSTTLload, TTLINPUTS (Pins
3, C,and 5) are
for use with transducers with high output levels greater than 2.4 V-DC. With 15-kΩ
impedance, LOW-LEVELINPUTS (Pins
4, D,and E) are 200 mV minimum and 9 V
maximum without damage, and are for use with transducers with high output levels of
200 mV-DC to 2.4 V-DC. Low transducer output is less than 0.8 V-DC for TTL, and less
than 50 mV for LOW-LEVEL.
GENERALDESCRIPTIONANDSPECIFICATIONSTRANSDUCERCONNECTIONS10A35.1
MODEL10A35
ENCODER
CONDITIONERCARD
SYSTEM10
ANALOGINPUTCARDS
1.E.2.10A35
Product specificaties
| Merk: | Daytronic |
| Categorie: | Niet gecategoriseerd |
| Model: | 10A35 |
Heb je hulp nodig?
Als je hulp nodig hebt met Daytronic 10A35 stel dan hieronder een vraag en andere gebruikers zullen je antwoorden
Handleiding Niet gecategoriseerd Daytronic
25 Januari 2024
25 Januari 2024
25 Januari 2024
25 Januari 2024
25 Januari 2024
25 Januari 2024
25 Januari 2024
25 Januari 2024
25 Januari 2024
25 Januari 2024
Handleiding Niet gecategoriseerd
Nieuwste handleidingen voor Niet gecategoriseerd
25 Augustus 2026
25 Augustus 2026
25 Augustus 2026
25 Augustus 2026
25 Augustus 2026
25 Augustus 2026
25 Augustus 2026
25 Augustus 2026
25 Augustus 2026
25 Augustus 2026